9 research outputs found

    Albumin determined by bromocresol green leads to erroneous results in routine evaluation of patients with chronic kidney disease

    Get PDF
    Objectives: Measurement of plasma albumin is pivotal for clinical decision-making in patients with chronic kidney disease (CKD). Routinely used methods as bromocresol green (BCG) and bromocresol purple (BCP) can suffer from aselectivity, but the impact of aselectivity on the accuracy of plasma albumin results of CKD-patients is still unknown. Therefore, we evaluated the performance of BCG-, BCP- and JCTLM-endorsed immunological methods in patients with various stages of CKD. Methods:We evaluated the performance of commonly used albumin methods in patients with CKD stages G1 through G5, the latter divided in two groups based on whether they received hemodialysis treatment. In total, 163 patient plasma samples were measured at 14 laboratories, on six different BCG and BCP-platforms, and four different immunological platforms. The results were compared with an ERM-DA-470k-corrected nephelometric assay. The implications on outcome is evaluated by the proportion of patient results &lt;38g/L for the diagnosis of protein energy wasting. Results:Albumin results determined with BCP- and immunological methods showed the best agreement with the target value (92.7 and 86.2%, respectively vs. 66.7% for BCG, namely due to overestimation). The relative agreement of each method with the target value was platform-dependent, with larger variability in agreement between platforms noted for BCG and immunological methods (3.2-4.6 and 2.6-5.3%) as opposed to BCP (0.7-1.5%). The stage of CKD had similar effects on the variability in agreement for the three method-groups (0.6-1.8% vs. 0.7-1.5% vs. 0.4-1.6%). The differences between methods cause discrepancies in clinical decision-making, as structurally fewer patients were diagnosed with protein energy wasting upon using BCG-based albumin results. Conclusions: Our study shows that BCP is fit for the intended use to measure plasma albumin levels in CKD patients from all stages, including patients on hemodialysis. In contrast, most BCG-based platforms falsely overestimate the plasma albumin concentration.</p

    Validation and verification of examination procedures in medical laboratories : opinion of the EFLM Working Group Accreditation and ISO/CEN standards (WG-A/ISO) on dealing with ISO 15189:2012 demands for method verification and validation

    No full text
    This paper reflects the opinion of the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group Accreditation and ISO/CEN standards (WG-A/ISO). It aims to provide guidance for drawing up local/national documents about validation and verification of laboratory methods. We demonstrate how risk evaluation can be used to optimize laboratory policies to meet intended use requirements as well as requirements of standards. This is translated in a number of recommendations on how to introduce risk evaluation in various stages of the implementation of new methods ultimately covering the whole process cycle

    A multicenter comparison of whole blood vitamin B6 assays

    No full text
    Background: The aim of this study was to compare different analytical methods that are currently in use in the Netherlands for the measurement of whole blood vitamin B6. Methods: This method comparison study consisted of two separate parts. (1) Four laboratories participated in a pilot study in which the commercial Chromsystems and INstruchemie method, and a laboratory developed liquid chromatography-tandem mass spectrometry (LC-MS/MS) method and HPLC method were compared. Sixty-nine frozen whole blood samples and six lyophilized whole blood samples were used for comparison. (2) In the nationwide part of the study, 49 laboratories participated in the analysis of three identical sets of two whole blood samples of which one set was freshly analyzed, one set was analyzed after storage at -20 degrees C and one set was analyzed after lyophilization. Results: In both parts of the study, the HPLC and LC-MS/MS methods showed equivalent results for all sample types tested. The Chromsystems method showed a positive bias of 45% (pilot study) and 30% (nationwide study) towards the LC-MS/MS method when fresh or frozen samples were used. The measurement of lyophilized samples showed no differences between the methods. The results of the INstruchemie method were inconclusive due to the low number of participants. Conclusions: The different analytical methods for measuring vitamin B6 produce different results when whole blood patient samples are measured. The recognition of a reference method or the development of suitable reference materials and quality control materials might serve as a first step towards improved standardization or harmonization of the whole blood vitamin B6 assay

    Albumin determined by bromocresol green leads to erroneous results in routine evaluation of patients with chronic kidney disease

    Get PDF
    Objectives: Measurement of plasma albumin is pivotal for clinical decision-making in patients with chronic kidney disease (CKD). Routinely used methods as bromocresol green (BCG) and bromocresol purple (BCP) can suffer from aselectivity, but the impact of aselectivity on the accuracy of plasma albumin results of CKD-patients is still unknown. Therefore, we evaluated the performance of BCG-, BCP- and JCTLM-endorsed immunological methods in patients with various stages of CKD.Methods: We evaluated the performance of commonly used albumin methods in patients with CKD stages G1 through G5, the latter divided in two groups based on whether they received hemodialysis treatment. In total, 163 patient plasma samples were measured at 14 laboratories, on six different BCG and BCP-platforms, and four different immunological platforms. The results were compared with an ERM-DA-470k-corrected nephelometric assay. The implications on outcome is evaluated by the proportion of patient results &lt;38 g/L for the diagnosis of protein energy wasting.Results: Albumin results determined with BCP- and immunological methods showed the best agreement with the target value (92.7 and 86.2 %, respectively vs. 66.7 % for BCG, namely due to overestimation). The relative agreement of each method with the target value was platform-dependent, with larger variability in agreement between platforms noted for BCG and immunological methods (3.2-4.6 and 2.6-5.3 %) as opposed to BCP (0.7-1.5 %). The stage of CKD had similar effects on the variability in agreement for the three method-groups (0.6-1.8 % vs. 0.7-1.5 % vs. 0.4-1.6 %). The differences between methods cause discrepancies in clinical decision-making, as structurally fewer patients were diagnosed with protein energy wasting upon using BCG-based albumin results.Conclusions: Our study shows that BCP is fit for the intended use to measure plasma albumin levels in CKD patients from all stages, including patients on hemodialysis. In contrast, most BCG-based platforms falsely overestimate the plasma albumin concentration.</p

    Production of galacto-oligosaccharides by the β-galactosidase from kluyveromyces lactis : comparative analysis of permeabilized cells versus soluble enzyme

    No full text
    8 páginas, 5 figuras, 3 tablas -- PAGS nros. 10477-10484The transgalactosylation activity of Kluyveromyces lactis cells was studied in detail. Cells were permeabilized with ethanol and further lyophilized to facilitate the transit of substrates and products. The resulting biocatalyst was assayed for the synthesis of galacto-oligosaccharides (GOS) and compared with two soluble β-galactosidases from K. lactis (Lactozym 3000 L HP G and Maxilact LGX 5000). Using 400 g/L lactose, the maximum GOS yield, measured by HPAEC-PAD analysis, was 177 g/L (44% w/w of total carbohydrates). The major products synthesized were the disaccharides 6-galactobiose [Gal-β(1→6)-Gal] and allolactose [Gal-β(1→6)-Glc], as well as the trisaccharide 6-galactosyl-lactose [Gal-β(1→6)-Gal-β(1→4)-Glc], which was characterized by MS and 2D NMR. Structural characterization of another synthesized disaccharide, Gal-β(1→3)-Glc, was carried out. GOS yield obtained with soluble β-galactosidases was slightly lower (160 g/L for Lactozym 3000 L HP G and 154 g/L for Maxilact LGX 5000); however, the typical profile with a maximum GOS concentration followed by partial hydrolysis of the newly formed oligosaccharides was not observed with the soluble enzymes. Results were correlated with the higher stability of β-galactosidase when permeabilized whole cells were usedProjects BIO2007-67708-C04-01, BIO2007-67708-C04-03, BIO2010-20508-C04-01, and BIO2010-20508-C04-04 from the Spanish Ministry of Science and Innovation supported this research. B.R.-C. and M.A.d.A. were supported by fellowships from the Spanish Ministries of Science and Innovation (FPI program) and Education and Culture (FPU program), respectivelyPeer reviewe