    Lesion Eccentricity Plays a Key Role in Determining the Pressure Gradient of Serial Stenotic Lesions:Results from a Computational Hemodynamics Study

    Get PDF
    Purpose: In arterial disease, the presence of two or more serial stenotic lesions is common. For mild lesions, it is difficult to predict whether their combined effect is hemodynamically significant. This study assessed the hemodynamic significance of idealized serial stenotic lesions by simulating their hemodynamic interaction in a computational flow model. Materials and Methods: Flow was simulated with SimVascular software in 34 serial lesions, using moderate (15 mL/s) and high (30 mL/s) flow rates. Combinations of one concentric and two eccentric lesions, all 50% area reduction, were designed with variations in interstenotic distance and in relative direction of eccentricity. Fluid and fluid–structure simulations were performed to quantify the combined pressure gradient. Results: At a moderate flow rate, the combined pressure gradient of two lesions ranged from 3.8 to 7.7 mmHg, which increased to a range of 12.5–24.3 mmHg for a high flow rate. Eccentricity caused an up to two-fold increase in pressure gradient relative to concentric lesions. At a high flow rate, the combined pressure gradient for serial eccentric lesions often exceeded the sum of the individual lesions. The relative direction of eccentricity altered the pressure gradient by 15–25%. The impact of flow pulsatility and wall deformability was minor. Conclusion: This flow simulation study revealed that lesion eccentricity is an adverse factor in the hemodynamic significance of isolated stenotic lesions and in serial stenotic lesions. Two 50% lesions that are individually non-significant can combine more often than thought to hemodynamic significance in hyperemic conditions. Graphical Abstract: (Figure presented.).</p

    The Supera Interwoven Nitinol Stent as a Flow Diverting Device in Popliteal Aneurysms

    Get PDF
    PURPOSE: The feasibility of using a compressed interwoven Supera stent as a flow diverting device for popliteal aneurysms was recently demonstrated in patients. It is unclear, however, what the optimal flow diverting strategy is, because of the fusiform shape of popliteal aneurysms and their exposure to triphasic flow. To assess this flow diverting strategy for popliteal aneurysms, flow profiles and thrombus formation likelihood were investigated in popliteal aneurysm models. MATERIALS AND METHODS: Six popliteal aneurysm models were created and integrated into a pulsatile flow set-up. These models covered a bent and a straight anatomy in three configurations: control, single-lined and dual-lined Supera stents. Two-dimensional flow velocities were visualized by laser particle image velocimetry. In addition, the efficacy of the stent configurations for promoting aneurysm thrombosis was assessed by simulations of residence time and platelet activation. RESULTS: On average for the two anatomies, the Supera stent led to a twofold reduction of velocities in the aneurysm for single-lined stents, and a fourfold reduction for dual-lined stents. Forward flow was optimally diverted, whereas backward flow was generally deflected into the aneurysm. The dual-lined configuration led to residence times of 15–20 s, compared to 5–15 s for the single stent configurations. Platelet activation potential was not increased by the flow diverting stents. CONCLUSION: A compressed Supera stent was successfully able to divert flow in a popliteal aneurysm phantom. A dual-lined configuration demonstrated superior hemodynamic characteristics compared to its single-lined counterpart. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00270-022-03118-x

    Visceral stent patency in fenestrated stent grafting for abdominal aortic aneurysm repair

    Get PDF
    ObjectiveFenestrated endovascular abdominal aortic aneurysm repair (F-EVAR) has been introduced for treatment of aneurysms in which visceral arteries are incorporated. Patency of target vessels has been reported to be excellent. Results of the use of stent grafts to accommodate visceral arteries in F-EVAR are presented in this study, including an overview of factors that affect outcome.MethodsAll patients treated with fenestrated stent grafts in a single center between November 2001 and October 2011 were reviewed. Patients treated for suprarenal, juxtarenal, and infrarenal short-necked aortic aneurysms were included. Patients with thoraco-abdominal aneurysms or aneurysms treated with grafts with fixed side branches were excluded. Polytetrafluoroethylene covered stents were used routinely since June 2005. Target vessels and stents were examined using computed tomography angiography reconstructions. Primary end points were primary patency, defined as the absence of occlusion, and loss of renal function. Secondary end points were technical success, stenosis (defined as a ≥50% angiographic diameter reduction), stent fracture, and mortality.ResultsA total of 138 patients with a median age of 73 years (range, 50-91 years) met the inclusion criteria. Median computed tomography angiography follow-up was 13 months (range, 1-97 months). In total, 392 target vessels were provided with 140 scallops and 252 fenestrations. Visceral stents (−grafts) were placed in 254 target vessels. Technical success was obtained in 249 arteries (98.0%). Overall stent patency of target vessels was 95.7% at 1 year and 88.6% at 4 years. Renal artery stent patency was 97.4% at 1 year and 91.2% at 4 years (96.8% and 89.1% for uncovered stents; 97.3% and 92.4% for covered stents, respectively). There was no significant difference in patency between covered and uncovered stents in renal arteries (P = .71). Renal artery stenosis occurred in 26 stented arteries (11.3%) and occlusion in seven arteries. Renal artery stent stenosis occurred significantly more in uncovered than in covered stents (P = .04). Stent fractures occurred more in uncovered than in covered stents (P = .01) and was associated with a significantly lower visceral stent patency rate (P < .01). During follow-up, 13 patients developed permanent renal function impairment (9.4%), of which two required permanent dialysis (1.4%). Renal dysfunction was significantly associated with renal stent occlusion or stenosis (P < .01).ConclusionsPatency rates of visceral artery stent (–grafts) in F-EVAR were 95.7% at 1 year and 88.6% at 4 years. Patency rates were affected by stent fractures, which occurred more in uncovered compared with covered stents. Renal artery stent stenosis occurred more in uncovered compared with covered stents. Renal dysfunction was significantly associated with renal stent occlusion or stenosis

    Eversion technique versus traditional carotid endarterectomy with patch angioplasty:a systematic review with meta-analyses and trial sequential analysis

    Get PDF
    Introduction: The use of an ‘eversion’ technique is not unequivocally proven to be superior to carotid endarterectomy with patch angioplasty. An up-to-date systematic review is needed for evaluation of benefits and harms of these two techniques. Methods: RCTs comparing eversion technique versus endarterectomy with patch angioplasty in patients with a symptomatic and significant (≥50 %) stenosis of the internal carotid artery were enrolled. Primary outcomes were all-cause mortality rate, health-related quality of life and serious adverse events. Secondary outcomes included 30-day stroke and mortality rate, (a) symptomatic arterial occlusion or restenosis, and adverse events not critical for decision making. Results: Four RCTs were included with 1272 surgical procedures for carotid stenosis; eversion technique n = 643 and carotid endarterectomy with patch closure n = 629. Meta-analysis comparing both techniques showed, with a very low certainty of evidence, that eversion technique might decrease the number of patients with serious adverse events (RR 0.47; 95% CI 0.34 to 0.64; p ≤ 0.01). However, no difference was found on the other outcomes. TSA demonstrated that the required information sizes were far from being reached for these patient-important outcomes. All patient-relevant outcomes were at low certainty of evidence according to GRADE. Conclusions: This systematic review showed no conclusive evidence of any difference between eversion technique and carotid endarterectomy with patch angioplasty in carotid surgery. These conclusions are based on data obtained in trials with very low certainty according to GRADE and should therefore be interpreted cautiously. Until conclusive evidence is obtained, the standard of care according to ESVS guidelines should not be abandoned.</p

    Incidence, natural course, and outcome of type II endoleaks in infrarenal endovascular aneurysm repair based on the ENGAGE registry data

    Get PDF
    Objective: The purpose of this study was to report the incidence, natural history, and outcome of type II endoleaks in the largest prospective real-world cohort to date. Methods: Patients were extracted from the prospective Endurant Stent Graft Natural Selection Global Postmarket Registry (ENGAGE). Two groups were analyzed: first, patients with an isolated type II endoleak; and second, patients with a type II endoleak who later presented with a type I endoleak. A health status analysis between patients with an early type II endoleak and patients with no endoleak was performed. Second, an attempt was made to identify risk factors in patients with a type II endoleak who later presented with a type I endoleak. Results: Through 5 years of follow-up, a total of 197 (15.6%) patients with isolated type II endoleaks were identified. Most were detected within the first 30 days (n = 73 [37.1%]) and through the first year (n = 73 [37.1%]), with the remainder being detected after 1 year of follow-up (n = 51 [25.8%]). Patients with a type II endoleak had a higher incidence of aneurysm growth and more secondary endovascular procedures (15.4% vs 7.5% at 5 years; P <.001). Overall survival was higher in the isolated type II endoleak group compared with patients with no endoleak (77.2% vs 67.0% at 5 years; P =.010). Twenty-two patients (10%) with a type II endoleak were diagnosed with a late type I endoleak (type IA, n = 10; type IB, n = 12), with a secondary intervention rate of 67.5% through 5 years. There was no difference in health status scores between patients with an early type II endoleak and patients without any type of endoleak at 1-year follow-up. Conclusions: In the ENGAGE registry, isolated type II endoleaks are present in 15.6% of patients during follow-up. The majority do not require secondary intervention, and an early isolated type II endoleak does not have an impact on health status through 1 year. However, a small group of patients with a type II endoleak will present with a type I endoleak, resulting in a high secondary intervention rate and significant risk of aneurysm-related complications

    Remodeling of Abdominal Aortic Angulation and Curvature After Endovascular Aneurysm Repair in Patients With vs Without Late Type Ia Endoleak or Endograft Migration

    Get PDF
    Purpose: To investigate aortic remodeling of the supra- and infrarenal aorta from preoperative to 1 month and midterm follow-up after endovascular aneurysm repair (EVAR) by analyzing changes in angulation and curvature in patients with vs without late type Ia endoleak or device migration. Materials and Methods: From a multicenter database, 35 patients (mean age 76 +/- 5 years; 31 men) were identified with late (>1 year) type Ia endoleak or endograft migration (>= 10 mm) and defined as the complication group. The control group consisted of 53 patients (mean age 75 +/- 7 years; 48 men) with >1-year computed tomography angiography (CTA) follow-up and no evidence of endoleaks. Suprarenal and infrarenal angles were measured on centerline reconstructions of the preoperative, 1-month, and midterm CTA scans. The value and location relative to baseline of maximum suprarenal and infrarenal curvature were determined semiautomatically using dedicated software. Changes were determined at 1 month compared with the preoperative CTA and at midterm compared with 1 month. Results: Preoperative suprarenal angulation was significantly greater in the complication group compared to the controls (34 degrees +/- 18 degrees vs 24 degrees +/- 17 degrees, p=0.008). It decreased significantly at 1 month in the complication group (29 degrees +/- 16 degrees, p=0.011) and at midterm follow-up in the controls (20 degrees +/- 19 degrees, p Conclusion: At midterm follow-up, significant differences in supra- and infrarenal angulation and curvature were observed between patients with vs without type Ia endoleak or migration. The location of the maximum curvature shifted distally in patients with complications. The aortic morphology is more stable during midterm follow-up in the patients without endoleaks

    Construct Validity and Reproducibility of Handheld Ultrasound Devices n Carotid Artery Diameter Measurement.

    Get PDF
    The construct validity and reproducibility of three commonly used handheld ultrasound (US) devices in measuring carotid arterial diameter was evaluated: Telemed MicrUs EXT-1H (Telemed, Vilnius, Lithuania), Butterfly iQ (Butterfly Network, Inc., Guilford, CT, USA) and Philips Lumify (Philips Healthcare, Best, The Netherlands). An in vitro setup was built to evaluate construct validity, compared with high-end US, and intra-observer variability of handheld US devices. Handheld devices showed a mean difference of 0.023 ± 0.030 cm, 0.012 ± 0.037 cm and 0.009 ± 0.046 cm for, respectively, Telemed, Butterfly and Lumify in comparison with high-end US devices. Intraclass agreement with the high-end system as well as intra-observer variability for handheld US devices was classified as excellent, with all values greater than 0.95. Subsequently, inter-observer variability of handheld US devices was investigated in an in vivo setup with 20 healthy volunteers. Inter-observer variability was classified as excellent for Telemed (0.901), good for Lumify (0.827) and moderate for Butterfly (0.684) with a difference of, respectively, 0.005 ± 0.031 cm, 0.020 ± 0.050 cm and -0.003 ± 0.033 cm. In conclusion, handheld US devices demonstrated an excellent construct validity and intra-observer variability. Additionally, excellent-to-good inter-observer variability for Telemed and Lumify was observed, and Butterfly demonstrated a moderate inter-observer agreement. These results indicate that handheld US devices are effective for measuring carotid arterial diameter
    • …