    Commissioning of a synchrotron-based proton beam therapy system for use with a Monte Carlo treatment planning system

    Get PDF
    This work tackles the commissioning and validation of a novel combination of a synchrotron-based proton beam therapy system (Hitachi, Ltd.) for use with a Monte Carlo treatment planning system (TPS). Four crucial aspects in this configuration have been investigated: (1) Monte Carlo-based correction performed by the TPS to the measured integrated depth-dose curves (IDD), (2) circular spot modelling with a single Gaussian function to characterize the synchrotron physical spot, which is elliptical, (3) the modelling of the range shifter that enables using only one set of measurements in open beams, and (4) the Monte Carlo dose calculation model in small fields. Integrated depth-dose curves were measured with a PTW Bragg peak chamber and corrected, with a Monte Carlo model, to account for energy absorbed outside the detector. The elliptical spot was measured by IBA Lynx scintillator, EBT3 films and PTW microDiamond. The accuracy of the TPS (RayStation, RaySearch Laboratories) at spot modelling with a circular Gaussian function was assessed. The beam model was validated using spread-out Bragg peak (SOBP) fields. We took single-point doses at several depths through the central axis using a PTW Farmer chamber, for fields between 2 × 2cm and 30 × 30cm. We checked the range-shifter modelling from open-beam data. We tested clinical cases with film and an ioni- zation chamber array (IBA Matrix). Sigma differences for spots fitted using 2D images and 1D profiles to elliptical and circular Gaussian models were below 0.22 mm. Differences between SOBP measurements at single points and TPS calculations for all fields between 5 × 5 and 30 × 30cm were below 2.3%. Smaller fields had larger differences: up to 3.8% in the 2 × 2cm field. Mean differences at several depths along the central axis were generally below 1%. Differences in range- shifter doses were below 2.4%. Gamma test (3%, 3 mm) results for clinical cases were generally above 95% for Matrix and film. Approaches for modelling synchrotron proton beams have been validated. Dose values for open and range- shifter fields demonstrate accurate Monte Carlo correction for IDDs. Elliptical spots can be successfully modelled using a circular Gaussian, which is accurate for patient calculations and can be used for small fields. A double-Gaussian spot can improve small-field calculations. The range-shifter modelling approach, which reduces clinical commissioning time, is adequat

    Hospital-based proton therapy implementation during the COVID pandemic: early clinical and research experience in a European academic institution

    Get PDF
    Introduction A rapid deploy of unexpected early impact of the COVID pandemic in Spain was described in 2020. Oncology practice was revised to facilitate decision-making regarding multimodal therapy for prevalent cancer types amenable to multidisciplinary treatment in which the radiotherapy component searched more efcient options in the setting of the COVID-19 pandemic, minimizing the risks to patients whilst aiming to guarantee cancer outcomes. Methods A novel Proton Beam Therapy (PBT), Unit activity was analyzed in the period of March 2020 to March 2021. Institutional urgent, strict and mandatory clinical care standards for early diagnosis and treatment of COVID-19 infection were stablished in the hospital following national health-authorities’ recommendations. The temporary trends of patients care and research projects proposals were registered. Results 3 out of 14 members of the professional staf involved in the PBR intra-hospital process had a positive test for COVID infection. Also, 4 out of 100 patients had positive tests before initiating PBT, and 7 out of 100 developed positive tests along the weekly mandatory special checkup performed during PBT to all patients. An update of clinical performance at the PBT Unit at CUN Madrid in the initial 500 patients treated with PBT in the period from March 2020 to November 2022 registers a distribution of 131 (26%) pediatric patients, 63 (12%) head and neck cancer and central nervous system neoplasms and 123 (24%) re-irradiation indications. In November 2022, the activity reached a plateau in terms of patients under treatment and the impact of COVID pandemic became sporadic and controlled by minor medical actions. At present, the clinical data are consistent with an academic practice prospectively (NCT05151952). Research projects and scientifc production was adapted to the pandemic evolution and its infuence upon professional time availability. Seven research projects based in public funding were activated in this period and preliminary data on molecular imaging guided proton therapy in brain tumors and post-irradiation patterns of blood biomarkers are reported. Conclusions Hospital-based PBT in European academic institutions was impacted by COVID-19 pandemic, although clinical and research activities were developed and sustained. In the post-pandemic era, the benefts of online learning will shape the future of proton therapy education

    Commissioning of a synchrotron-based proton beam therapy system for use with a Monte Carlo treatment planning system

    No full text
    This work tackles the commissioning and validation of a novel combination of a synchrotron-based proton beam therapy system (Hitachi, Ltd.) for use with a Monte Carlo treatment planning system (TPS). Four crucial aspects in this configuration have been investigated: (1) Monte Carlo-based correction performed by the TPS to the measured integrated depth-dose curves (IDD), (2) circular spot modelling with a single Gaussian function to characterize the synchrotron physical spot, which is elliptical, (3) the modelling of the range shifter that enables using only one set of measurements in open beams, and (4) the Monte Carlo dose calculation model in small fields. Integrated depth-dose curves were measured with a PTW Bragg peak chamber and corrected, with a Monte Carlo model, to account for energy absorbed outside the detector. The elliptical spot was measured by IBA Lynx scintillator, EBT3 films and PTW microDiamond. The accuracy of the TPS (RayStation, RaySearch Laboratories) at spot modelling with a circular Gaussian function was assessed. The beam model was validated using spread-out Bragg peak (SOBP) fields. We took single-point doses at several depths through the central axis using a PTW Farmer chamber, for fields between 2 × 2cm and 30 × 30cm. We checked the range-shifter modelling from open-beam data. We tested clinical cases with film and an ioni- zation chamber array (IBA Matrix). Sigma differences for spots fitted using 2D images and 1D profiles to elliptical and circular Gaussian models were below 0.22 mm. Differences between SOBP measurements at single points and TPS calculations for all fields between 5 × 5 and 30 × 30cm were below 2.3%. Smaller fields had larger differences: up to 3.8% in the 2 × 2cm field. Mean differences at several depths along the central axis were generally below 1%. Differences in range- shifter doses were below 2.4%. Gamma test (3%, 3 mm) results for clinical cases were generally above 95% for Matrix and film. Approaches for modelling synchrotron proton beams have been validated. Dose values for open and range- shifter fields demonstrate accurate Monte Carlo correction for IDDs. Elliptical spots can be successfully modelled using a circular Gaussian, which is accurate for patient calculations and can be used for small fields. A double-Gaussian spot can improve small-field calculations. The range-shifter modelling approach, which reduces clinical commissioning time, is adequat

    Practice-oriented solutions integrating intraoperative electron irradiation and personalized proton therapy for recurrent or unresectable cancers: Proof of concept and potential for dual FLASH effect

    No full text
    Background: Oligo-recurrent disease has a consolidated evidence of long-term surviving patients due to the use of intense local cancer therapy. The latter combines real-time surgical exploration/resection with high-energy electron beam single dose of irradiation. This results in a very precise radiation dose deposit, which is an essential element of contemporary multidisciplinary individualized oncology. Methods: Patient candidates to proton therapy were evaluated in Multidisciplinary Tumor Board to consider improved treatment options based on the institutional resources and expertise. Proton therapy was delivered by a synchrotron-based pencil beam scanning technology with energy levels from 70.2 to 228.7 MeV, whereas intraoperative electrons were generated in a miniaturized linear accelerator with dose rates ranging from 22 to 36 Gy/min (at Dmax) and energies from 6 to 12 MeV. Results: In a period of 24 months, 327 patients were treated with proton therapy: 218 were adults, 97 had recurrent cancer, and 54 required re-irradiation. The specific radiation modalities selected in five cases included an integral strategy to optimize the local disease management by the combination of surgery, intraoperative electron boost, and external pencil beam proton therapy as components of the radiotherapy management. Recurrent cancer was present in four cases (cervix, sarcoma, melanoma, and rectum), and one patient had a primary unresectable locally advanced pancreatic adenocarcinoma. In re-irradiated patients (cervix and rectum), a tentative radical total dose was achieved by integrating beams of electrons (ranging from 10- to 20-Gy single dose) and protons (30 to 54-Gy Relative Biological Effectiveness (RBE), in 10–25 fractions). Conclusions: Individual case solution strategies combining intraoperative electron radiation therapy and proton therapy for patients with oligo-recurrent or unresectable localized cancer are feasible. The potential of this combination can be clinically explored with electron and proton FLASH beams

    Hospital-based proton therapy implementation during the COVID pandemic: early clinical and research experience in a European academic institution

    No full text
    Introduction A rapid deploy of unexpected early impact of the COVID pandemic in Spain was described in 2020. Oncology practice was revised to facilitate decision-making regarding multimodal therapy for prevalent cancer types amenable to multidisciplinary treatment in which the radiotherapy component searched more efcient options in the setting of the COVID-19 pandemic, minimizing the risks to patients whilst aiming to guarantee cancer outcomes. Methods A novel Proton Beam Therapy (PBT), Unit activity was analyzed in the period of March 2020 to March 2021. Institutional urgent, strict and mandatory clinical care standards for early diagnosis and treatment of COVID-19 infection were stablished in the hospital following national health-authorities’ recommendations. The temporary trends of patients care and research projects proposals were registered. Results 3 out of 14 members of the professional staf involved in the PBR intra-hospital process had a positive test for COVID infection. Also, 4 out of 100 patients had positive tests before initiating PBT, and 7 out of 100 developed positive tests along the weekly mandatory special checkup performed during PBT to all patients. An update of clinical performance at the PBT Unit at CUN Madrid in the initial 500 patients treated with PBT in the period from March 2020 to November 2022 registers a distribution of 131 (26%) pediatric patients, 63 (12%) head and neck cancer and central nervous system neoplasms and 123 (24%) re-irradiation indications. In November 2022, the activity reached a plateau in terms of patients under treatment and the impact of COVID pandemic became sporadic and controlled by minor medical actions. At present, the clinical data are consistent with an academic practice prospectively (NCT05151952). Research projects and scientifc production was adapted to the pandemic evolution and its infuence upon professional time availability. Seven research projects based in public funding were activated in this period and preliminary data on molecular imaging guided proton therapy in brain tumors and post-irradiation patterns of blood biomarkers are reported. Conclusions Hospital-based PBT in European academic institutions was impacted by COVID-19 pandemic, although clinical and research activities were developed and sustained. In the post-pandemic era, the benefts of online learning will shape the future of proton therapy education

    SRC family kinase (SFK) inhibitor dasatinib improves the antitumor activity of anti-PD-1 in NSCLC models by inhibiting Treg cell conversion and proliferation

    Get PDF
    Introduction The use of immune-checkpoint inhibitors has drastically improved the management of patients with non-small cell lung cancer (NSCLC), but innate and acquired resistances are hurdles needed to be solved. Immunomodulatory drugs that can reinvigorate the immune cytotoxic activity, in combination with antiprogrammed cell death 1 (PD-1) antibody, are a great promise to overcome resistance. We evaluated the impact of the SRC family kinases (SFKs) on NSCLC prognosis, and the immunomodulatory effect of the SFK inhibitor dasatinib, in combination with anti-PD-1, in clinically relevant mouse models of NSCLC. Methods A cohort of patients from University Clinic of Navarra (n=116) was used to study immune infiltrates by multiplex immunofluorescence (mIF) and YES1 protein expression in tumor samples. Publicly available resources (TCGA, Km Plotter, and CIBERSORT) were used to study patient's survival based on expression of SFKs and tumor infiltrates. Syngeneic NSCLC mouse models 393P and UNSCC680AJ were used for in vivo drug testing. Results Among the SFK members, YES1 expression showed the highest association with poor prognosis. Patients with high YES1 tumor levels also showed high infiltration of CD4+/FOXP3+ cells (regulatory T cells (Tregs)), suggesting an immunosuppressive phenotype. After testing for YES1 expression in a panel of murine cell lines, 393P and UNSCC680AJ were selected for in vivo studies. In the 393P model, dasatinib+anti-PD-1 treatment resulted in synergistic activity, with 87% tumor regressions and development of immunological memory that impeded tumor growth when mice were rechallenged. In vivo depletion experiments further showed that CD8+ and CD4+ cells are necessary for the therapeutic effect of the combination. The antitumor activity was accompanied by a very significant decrease in the number of Tregs, which was validated by mIF in tumor sections. In the UNSCC680AJ model, the antitumor effects of dasatinib+anti-PD-1 were milder but similar to the 393P model. In in vitro assays, we demonstrated that dasatinib blocks proliferation and transforming growth factor beta-driven conversion of effector CD4+ cells into Tregs through targeting of phospholymphocyte-specific protein tyrosine kinase and downstream effectors pSTAT5 and pSMAD3. Conclusions YES1 protein expression is associated with increased numbers of Tregs in patients with NSCLC. Dasatinib synergizes with anti-PD-1 to impair tumor growth in NSCLC experimental models. This study provides the preclinical rationale for the combined use of dasatinib and PD-1/programmed death-ligand 1 blockade to improve outcomes of patients with NSCLC

    Adolescent's physical activity levels and relatives' physical activity engagement and encouragement: The HELENA study

    No full text

    Cultural differences in positive psychotic experiences assessed with the Community Assessment of Psychic Experiences-42 (CAPE-42):a comparison of student populations in the Netherlands, Nigeria and Norway

    No full text
    BACKGROUND: Previous studies have suggested that culture impacts the experience of psychosis. The current study set out to extend these findings by examining cultural variation in subclinical positive psychotic experiences in students from The Netherlands, Nigeria, and Norway. Positive psychotic experiences were hypothesized to (i) be more frequently endorsed by, and (ii) cause less distress in Nigerian vs. Dutch and Norwegian students. METHODS: Psychology students, aged 18 to 30 years, from universities in the Netherlands (n = 245), Nigeria (n = 478), and Norway (n = 162) were assessed cross-sectionally with regard to the frequency of subclinical positive psychotic experiences and related distress, using the Community Assessment of Psychic Experiences (CAPE-42). Multi-group confirmatory factor analysis and multivariate analysis of covariance were performed to assess measurement invariance of the positive symptom dimension (CAPE-Pos) and compare mean frequency and associated distress of positive psychotic experiences across study samples. RESULTS: Only CAPE-Pos items pertaining to the dimensions 'strange experiences' and 'paranoia' met assumptions for (partial) measurement invariance. Frequencies of these experiences were higher in the Nigerian sample, compared to both the Dutch and Norwegian samples, which were similar. In addition, levels of experience-related distress were similar or higher in the Nigerian sample compared to respectively the Dutch and Norwegian samples. CONCLUSION: Although positive psychotic experiences may be more commonly endorsed in non-Western societies, our findings do not support the notion that they represent a more benign, and hence less distressing aspect of human experience. Rather, the experience of psychotic phenomena may be just as, if not more, distressing in African than in European culture. However, observed differences in CAPE-Pos frequency and distress between samples from different cultural settings may partly reflect differences in the measure rather than in the latent trait. Future studies may therefore consider further cross-cultural adaptation of CAPE-42, in addition to explicitly examining cultural acceptance of psychotic phenomena, and environmental and other known risk factors for psychosis, when comparing and interpreting subclinical psychotic phenomena across cultural groups