    Results of the EURAMET.RI(II)- S7.Sm-151 supplementary comparison (EURAMET Project 1292)

    No full text
    An international comparison of the activity standardisation of the relatively long-lived gamma-ray emitter 151Sm has been recently completed. A total of six laboratories measured a solution prepared by CEA/LNHB and CEA/LANIE. Aliquots of the master solution were standardized in terms of activity per mass unit by participant laboratories using 2 different techniques. The results of the comparison can be used as the basis for establishing equivalence among the laboratories. The activity measurements of this comparison are part of the joint research project “Metrology for Radioactive Waste Management” of the European Metrology Research Programme (EMRP). One aim of this project is a new determination of the 151Sm half-life.JRC.D.4-Standards for Nuclear Safety, Security and Safeguard

    Determination of the 151Sm half-life

    No full text
    New measurements have been undertaken to determine the half-life of 151Sm. A pure 151Sm solution was obtained after chemical separation from a samarium solution resulting from the dissolution of an irradiated samarium sample. The concentration of 151Sm in the solution was measured by mass spectrometry, combined with the isotope dilution technique. The activity of the solution was measured by liquid scintillation counting by six European laboratories as part of an international comparison. These combined results lead to a half-life of T1/2 = 94.6(6) a.JRC.D.4-Standards for Nuclear Safety, Security and Safeguard