12 research outputs found

    The intraocular electrode array of the suprachoroidal device (A) and the entire device (B), showing the array connected to the percutaneous connector via a helical lead wire.

    No full text
    <p>The electrodes on the intraocular array (C) were numbered for analysis, with the black electrodes (21a to 21m) being ganged to provide an external ring for common ground and hexagonal stimulation parameter testing. Note electrodes 9, 17 and 19 were smaller (400 µm vs. 600 µm). The percutaneous connector protruded through the skin behind the ear (D), allowing direct connection to the neurostimulator via a connecting lead (E). The scleral incision was made 9 mm to 10 mm posteriorly from the sclero-corneal limbus.</p

    Impedances for the 600 µm platinum electrodes over time in the three subjects.

    No full text
    <p>Impedances were measured with charge-balanced biphasic current pulses (pulse phase width: 25 µs; amplitude: 75 µA). The dotted lines represent the date of first stimulation. In P1 & P2, the impedances were stable over the implantation and stimulation period. Impedances measured in P3 decreased over the course of the implantation period. Outliers are identified by open circles, and the numbers represent the electrode location (see <a href="http://www.plosone.org/article/info:doi/10.1371/journal.pone.0115239#pone-0115239-g002" target="_blank">Fig. 2</a>).</p

    Distance between the electrodes and retina over time.

    No full text
    <p>This distance was relatively constant in P1, but increased up to two-fold in P2 and P3 over time. Note significant nystagmus in P3 made the measurements difficult, leading to greater variation in the values recorded. Outliers are identified by open circles, and the numbers represent the electrode location (see <a href="http://www.plosone.org/article/info:doi/10.1371/journal.pone.0115239#pone-0115239-g002" target="_blank">Fig. 2</a>).</p

    Retinal thickness measurements over time, showing no observable change in the maximum retinal thickness above the electrode array in the initial twelve months in all three patients.

    No full text
    <p>Each boxplot includes the maximum (upper whisker, excluding outliers), upper quartile (top of box), median (horizontal line in box), lower quartile (bottom of box) and minimum values (lower whisker). Open circles are outliers. The numbers represent the electrode location (see <a href="http://www.plosone.org/article/info:doi/10.1371/journal.pone.0115239#pone-0115239-g002" target="_blank">Fig. 2</a>), and the horizontal lines on the graph of P3 represent single data points.</p