    Premature ventricular contractions in patients with an implantable cardioverter defibrillator cardiac resynchronization therapy device: Results from the UMBRELLA registry

    Get PDF
    BACKGROUND: Premature ventricular contractions (PVC) are known to reduce the percentage of biventricular (BiV) pacing in patients with cardiac resynchronization (CRT), decreasing the clinical response. The aim of this study was to evaluate the prevalence of a high PVC burden, as well as therapeutic action (pharmacotherapy, catheter ablation or device programming), in a large CRT implantable-defibrillator (CRT-D) population. METHODS: Patients with a CRT-D device from the UMBRELLA multicenter prospective remote monitoring registry were included. The PVC count was collected from each remote monitoring transmission. Patients were divided into two high (>/=1 transmission >/=200/>/=400 PVC/h, respectively) and one low (all transmissions /=200/>/=400 PVC/h, respectively). The majority of patients in the high PVC groups were not treated (61 [79%] and 32 [74%], respectively. Considering the untreated patients in the high PVC groups, median PVC/h was 199 (interquartile range [IQR]: 196) and 271 (IQR: 330), respectively. The PVC burden (proportion of time with PVC/h >/= 200/>/=400) was 40% (IQR 70) and 29% (IQR 59), respectively. CONCLUSION: A significant proportion of CRT-D patients presented a high PVC count, however, few received treatment. In the untreated patients with a high PVC count, the PVC burden during follow-up varied substantially. Several consecutive recordings of a high PVC count should be warranted before considering therapeutic action such as catheter ablation