45 research outputs found

    Impact of percutaneous mitral valve repair using the MitraClipℱ system on ventricular arrhythmias and ICD therapies

    Get PDF
    Transcatheter edge-to-edge repair (TEER) using the MitraClipℱ device has been established as a suitable alternative to mitral valve surgery in patients with severe mitral regurgitation (MR) and high or prohibitive surgical risk. Only limited information regarding the impact of TEER on ventricular arrhythmias (VA) has been reported. The aim of the present study was to assess the impact of TEER using the MitraClip(TM) device on the burden of VA and ICD (Implantable Cardioverter Defibrillator) therapies. Among 600 MitraClip(TM) implantations performed in our clinic between September 2009 and October 2018, we identified 86 patients with successful TEER and an active implantable cardiac device (pacemaker, ICD, CRT-P/D (Cardiac Resynchronization Therapy-Pacemaker/Defibrillator)) eligible for retrospective VA analyses. These patients presented with mainly functional MR (81.4%) and severely reduced left ventricular ejection fraction (mean LVEF 22.1% ± 10.3%). The observation period comprised 456 ± 313 days before and 424 ± 287 days after TEER. The burden of ventricular arrhythmias (sustained ventricular tachycardia (sVT) and ventricular fibrillation (VF)) was significantly reduced after TEER (0.85 ± 3.47 vs. 0.43 ± 2.03 events per patient per month, p = 0.01). Furthermore, the rate of ICD therapies (anti-tachycardia pacing (ATP) and ICD shock) decreased significantly after MitraClip(TM) implantation (1.0 ± 3.87 vs. 0.32 ± 1.41, p = 0.014). However, reduction of VA burden did not result in improved two-year survival in this patient cohort with severely reduced LVEF. Mitral valve TEER using the MitraClipℱ device was associated with a significant reduction of ventricular arrhythmias and ICD therapies

    Re‐do MitraClip in patients with functional mitral valve regurgitation and advanced heart failure

    Get PDF
    AIM: Percutaneous mitral valve repair (PMVR) via MitraClip implantation is a therapeutic option for severe mitral regurgitation (MR) in advanced stages of heart failure (HF). However, progressive left ventricular dilation in these patients may lead to recurrent MR after PMVR and consequent re‐do MitraClip implantation. Here, we describe the characteristics and outcomes of this clinical scenario. METHODS AND RESULTS: Patients with systolic HF and functional MR undergoing a re‐do MitraClip procedure were retrospectively analysed. Inclusion criteria were age ≄18 years, technical, device and procedural success at first MitraClip procedure, functional MR and systolic HF with an ejection fraction (EF) of <45%. Seventeen out of 684 patients undergoing PMVR with the MitraClip device at our institution between September 2009 and July 2019 were included. All patients displayed advanced HF with an EF of 20% (±9.9) and highly elevated N‐terminal pro‐brain natriuretic peptide. Technical success of the re‐do MitraClip procedure was 100%, whereas procedural and device success were only achieved in 11 patients (65%). Unsuccessful re‐do procedures were related to lower EF and implantation of more than one clip at initial procedure. However, despite reduction in MR grade and no occurrence of significant mitral stenosis after the procedure, the mortality during 12 months follow‐up remained high (8 of 17; 47%). CONCLUSIONS: In a cohort of patients with advanced HF undergoing PMVR, re‐do MitraClip procedure was feasible, but procedural success was unsatisfactory and morbidity and mortality remained high, possibly reflecting the advanced stage of HF in these patients

    Abfall von SauerstoffsÀttigung und Blutdruck sowie Anstieg des zentralen Venendrucks im Rahmen eines Mitralklappenclippings bei einer 81-JÀhrigen

    Get PDF
    BACKGROUND: Atrial septal defects (ASD) following endovascular mitral valve clipping are potentially hemodynamically relevant complications. Immediate closure with an occluder can represent a safe and effective treatment. CASE SUMMARY: An 81-year-old female patient suffering from severe dyspnea due to previously known severe mitral valve regurgitation was scheduled for elective mitral valve clipping. The clip was successfully implanted. Removal of the transseptal cannula resulted in a sudden drop in oxygen saturation and systolic blood pressure as well as an immediate increase in central venous pressure. An iatrogenic left-right shunt was observed at the atrial level with a relevant shunt volume. Immediate closure using an atrial septal occluder successfully restored the oxygen saturation and hemodynamic parameters. CONCLUSION: An increase in central venous pressure, reduction of systolic blood pressure or oxygen saturation after withdrawal of the transseptal cannula during mitral valve clipping should always be further investigated regarding a possible ASD
    corecore