37 research outputs found

    Metal-on-metal total hip arthroplasty: does increasing modularity affect clinical outcome?

    Get PDF
    BACKGROUND: Modularity of metal-on-metal (MoM) implants has come under scrutiny due to concerns regarding additional sources of metal debris. This study is a retrieval analysis of implants from the same manufacturer with the same MoM bearing surface. The difference between the implants was presence or absence of modular junctions. METHODS: This is a retrospective study of 31 retrieved implants from 31 patients who received a Conserve Wright Medical MoM hip prosthesis. The 31 implants consisted of 16 resurfacings and 15 implants with modular junctions; 4 conventional THAs and 11 modular-neck THAs. RESULTS: 43% of pre-revision MRI scans performed on resurfacing implants and 91% performed on the modular implants illustrated evidence of an adverse local tissue reaction. There was no difference in pre-revision blood metal ion levels or bearing surface wear between the resurfacings and modular implants. The neck-head tapers of the modular group showed low levels of material loss. However, the neck-stem tapers showed increased severity of corrosion and material loss. CONCLUSIONS: The modular implants had an increased incidence of adverse local tissue reaction. This could be related to the presence of modular junctions, particular the neck-stem junction which showed increased susceptibly to corrosion

    Does modularity of metal-on-metal hip implants increase cobalt: chromium ratio?

    Get PDF
    Background: Blood metal ion levels are used in the surveillance of metal-on-metal (MoM) hip implants. Modular implants contain an extra source of metal debris that may affect the ratio of metal ions in the blood. Methods: This was a retrospective study of 503 patients with hip replacements made by a single manufacturer (Smith & Nephew, Warwick, UK) with the same bearing surface. There were 54 total hip arthroplasties, 35 Birmingham Mid-Head Resections and 414 hip resurfacings. Whole blood metal ion levels and their ratios were analysed to investigate the effect of a modular junction. Results: The cobalt:chromium ratios were greater in the total hip arthroplasty group (mean 2.3:1) when compared to the resurfacings group (mean 1.3:1, p = <0.05) and Birmingham Mid-Head Resection group (mean 1.1:1, p = 0.11). Conclusions: This study demonstrated a trend for a higher cobalt:chromium ratio in patients with MoM total hip replacement that may be due to metal debris from the modular stem-head junction. Further work is required to correlate clinical data with retrieval analysis to confirm the effect of taper material loss on the cobalt:chromium ratio

    Variation in taper surface roughness for a single design effects the wear rate in total hip arthroplasty

    Get PDF
    Material loss from the head-stem taper junction of total hip arthroplasty (THA) is implicated in adverse reactions to metal debris (ARMD); the mechanisms for this are multi-factorial. We investigated the relationship between the roughness of the "as manufactured" taper surface and the wear rate from this junction. Fifty retrieved Pinnacle metal-on-metal (MOM) bearings paired with a Corail stem were included in the study. Multivariable statistical analysis was performed to determine the influence of taper roughness on material loss rate after controlling for other confounding surgical, implant, and patient factors. The surface roughness of the "as manufactured" head taper surface was associated with the rate of material loss from this surface. Four of eighteen roughness variables taken from ISO 4,287 and ISO 13,565-2 were significant: The Reduced Peak Height (Rpk, the protruding peaks above the core) (p = 0.004), Material Ratio 1 (Mr1, the ratio of the protruding peaks above the core) (p = 0.002), Area of the Peak Region (A1, the area of the Abbott-Curve that contains the peaks from the profile) (p = 0.003) and the Skewness (Rsk, the asymmetry of the height distribution corresponding to the height or depth of surface features) (p = 0.03). We found a large variability in the measured values with a median (range) of 0.50 (0.05-2.98), 11.98 (0.46-39.98), 30.89 (0.15-581.00), and 0.04 (-0.73-0.84), respectively. A 1-unit increase in Rpk was associated with a 73% increase in the taper wear rate. The variability of "as manufactured" surface roughness has a significant effect on taper material loss. © 2016 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res

    Clinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty.

    Get PDF
    OBJECTIVES: Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. METHODS: In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons. RESULTS: Evidence of corrosion was observed in 55% of hips. The median Goldberg taper corrosion score was 2 (1 to 4) and the annual rate of material loss at the taper was 0.084 mm(3)/year (0 to 0.239). The median trunnion corrosion score was 1 (1 to 3). CONCLUSIONS: We have reported a level of trunnionosis for MOP hips with large-diameter heads that were revised for reasons other than trunnionosis, and therefore may be clinically insignificant.Cite this article: H. S. Hothi, D. Kendoff, C. Lausmann, J. Henckel, T. Gehrke, J. Skinner, A. Hart. Clinically insignificant trunnionosis in large-diameter metal-on-polyethylene total hip arthroplasty. Bone Joint Res 2017;6:52-56. DOI: 10.1302/2046-3758.61.BJR-2016-0150.R2

    Clinical Cold Welding of the Modular Total Hip Arthroplasty Prosthesis.

    Get PDF
    BACKGROUND: A head that is "clinically cold welded" to a stem is one of the commonest reasons for unplanned removal of the stem. It is not clear which hip designs are at greatest risk of clinical cold welding. METHODS: This was a case-control study of consecutively received hip implant retrievals; we chose the design of hip that had the greatest number of truly cold-welded heads (n = 11). For our controls, we chose retrieved hips of the same design but without cold welding of the head (n = 35). We compared the clinical variables between these 2 groups using nonparametric Mann-Whitney tests to investigate the significance of differences between the cold-welded and non-cold-welded groups. RESULTS: The design that most commonly caused cold welding was a combination of a Ti stem and Ti taper: 11 out of 48 (23%) were truly cold welded. Comparison of the clinical data showed that no individual factor could be used to predict this preoperatively with none of the 4 predictors tested showing any significance: (1) time to revision (P = .687), (2) head size (P = .067), (3) patient age at primary (P = .380), and (4) gender (P = .054). CONCLUSION: We have shown that clinical cold welding is most prevalent in Ti-Ti combinations of the stem and taper; approximately 25% of cases received at our center were cold welded. Analysis of clinical variables showed that it is not possible to predict which will be cold welded preoperatively. Surgeons should be aware of this potential complication when revising a Ti-Ti stem/head junction

    Factors Associated With Trunnionosis in the Metal-on-Metal Pinnacle Hip

    Get PDF
    BACKGROUND: Trunnionosis of the tapered head-stem junction of total hip arthroplasties, either through corrosion or mechanical wear, has been implicated in early implant failure. Retrieval analysis of large numbers of failed implants can help us better understand the factors that influence damage at this interface. METHODS: In this study, we examined 120 retrieved total hip arthroplasties of one bearing design, the 36-mm diameter metal-on-metal, DePuy Pinnacle, that had been paired with 3 different stems. We measured material loss of the bearing and head-trunnion taper surfaces and collected clinical and component data for each case. We then used multiple linear regression analysis to determine which factors influenced the rate of taper material loss. RESULTS: We found 4 significant variables: (1) longer time to revision (P = .004), (2) the use of a 12/14 taper for the head-trunnion junction (P < .001), (3) decreased bearing surface wear (P = .003), and (4) vertical femoral offset (P = .05). These together explained 29% of the variability in taper material loss. CONCLUSION: Our most important finding is the effect of trunnion design. Of the 3 types studied, we found that S-ROM design was the most successful at minimizing trunnionosis

    Retrieval evidence of impingement at the third articulation in contemporary dual mobility cups for total hip arthroplasty

    Get PDF
    PURPOSE: We aimed to assess polyethylene liners of retrieved hips of one design of a dual mobility (DM) cup liner and two designs of femoral stems to better understand the role of femoral stem design on polyethylene impingement. METHODS: This was a case-control study involving 70 retrieved highly cross-linked polyethylene (X3) liners used with ABGII (n = 35) and Rejuvenate (n = 35) stems (Stryker). All polyethylene liners were assessed for evidence of rim deformation and the damage quantified using metrology methods. RESULTS: A total of 80% of polyethylene liners paired with ABGII necks had macroscopic evidence of neck impingement resulting in a raised lip whilst 23% of liners paired with Rejuvenate necks had evidence of a raised lip (p < 0.0001). The height of the raised rims of the DM cups paired with ABGII necks had a median (range) of 139 μm (72-255). The height of the raised rims of the DM cups paired with Rejuvenate necks had a median (range) of 52 μm (45-90) (p < 0.0001). CONCLUSION: Our new findings from retrieved dual mobility bearings showed that polyethylene liner rim deformation resulting from impingement with the femoral neck occurs in early in-human function, is circumferential in distribution, and is affected by the stem neck design. We recommend the use of highly polished and non-edged neck designs when used in conjunction with DM cups

    Mechanical wear analysis helps understand a mechanism of failure in retrieved magnetically controlled growing rods: a retrieval study

    Get PDF
    BACKGROUND: To assess the relationship between mechanical wear and the failure of the internal lengthening mechanism in retrieved MAGnetic Expansion Control (MAGEC) growing rods. METHODS: This study included 34 MAGEC rods retrieved from 20 patients. The state of the internal mechanism and mechanical wear were assessed in all the rods using plain radiographs and visual inspection. Metrology was then performed to assess the topography and mechanical wear of the telescopic bars, using a Talyrond 365 (Taylor Hobson, Leicester, UK) roundness measuring machine. RESULTS: Plain radiographs showed evidence of a broken internal mechanism in 29% of retrieved rods. Single-side wear marks were found in 97% of retrieved rods. Material loss was found to significantly increase in rods with a damaged internal mechanism (p < 0.05) and rods with longer time in situ (r = 0.692, p < 0.05). CONCLUSIONS: We found an association between damage to the internal mechanism of the rods and (1) patterns of single-side longitudinal wear marks and (2) increased material loss. As the material loss was also found to increase over time of rod in situ, we emphasise the importance of early detection and revision of failed MAGEC rods in clinical practice

    Analysing a mechanism of failure in retrieved magnetically controlled spinal rods

    Get PDF
    PURPOSE: We aim to describe a mechanism of failure in magnetically controlled growth rods which are used for the correction of the early onset scoliosis. METHODS: This retrieval study involved nine magnetically controlled growth rods, of a single design, revised from five patients for metal staining, progression of scoliosis, swelling, fractured actuator pin, and final fusion. All the retrieved rods were radiographed and assessed macroscopically and microscopically for material loss. Two implants were further analysed using micro-CT scanning and then sectioned to allow examination of the internal mechanism. No funding was obtained to analyse these implants. There were no potential conflicts interests. RESULTS: Plain radiographs revealed that three out of nine retrieved rods had a fractured pin. All had evidence of surface degradation on the extendable telescopic rod. There was considerable corrosion along the internal mechanism. CONCLUSIONS: We found that a third of the retrieved magnetically controlled growth rods had failed due to pin fracture secondary to corrosion of the internal mechanism. We recommend that surgeons consider that any inability of magnetically controlled growth rods to distract may be due to corrosive debris building up inside the mechanism, thereby preventing normal function

    Wear performance of retrieved metal-on-metal Pinnacle hip arthroplasties implanted before and after 2007

    Get PDF
    Objectives: Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. Methods: A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The 'pre-2007' group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the '2007 onwards' group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors. Results: There was no significant difference between the taper and bearing wear rates of the 'pre-2007' and '2007 onwards' groups (p = 0.67 and p = 0.39, respectively). Pinnacles implanted from 2007 onwards were revised after a mean time of 50 months, which was significantly earlier than the 'pre-2007' hips (96 months) (p < 0.001). A reduction in the time to revision was present year on year from 2003 to 2011. Conclusion: We found no difference in the wear rate of these implants based on the year of implantation. The 'pre-2007' hips had a two-fold greater time to revision than those implanted after 2007; this may be due to the increased surveillance of MoM hips following UK regulatory advice and several high-profile failures. Interestingly, we observed a decreasing trend in the mean time to revision every year from 2003 onwards.Cite this article: S. Bergiers, H. S. Hothi, J. Henckel, A. Eskelinen, J. Skinner, A. Hart. Wear performance of retrieved metal-on-metal Pinnacle hip arthroplasties implanted before and after 2007