19 research outputs found

    Failure of dual radius hydroxyapatite-coated acetabular cups

    Get PDF
    <p>Abstract</p> <p>Introduction</p> <p>Many kind of hydroxyapatite-coated cups were used, with favorable results in short term studies; it was supposed that its use could improve osteointegration of the cup, enhancing thus stability and survivorship. The purpose of this study is to analyze the long term behavior of the hemispheric HA coated, Dual Radius Osteonics cup and to discuss the way of failure through the exam of the revised components and of both periacetabular and osteolysis tissue.</p> <p>Materials and Methods</p> <p>Between 1994 and 1997, at the Department of Orthopedic Sciences of the Insubria University, using the posterolateral approach, were implanted 276 Dual Radius Osteonics<sup>¼ </sup>in 256 patients, with mean age of 63 years.</p> <p>Results</p> <p>At a mean follow-up of 10 years (range 8–12 years), 183 cups in 165 patients, were available for clinical and radiographical evaluation. 22 Cups among the 183 were revised (11%). The cause of revision was aseptic loosening in 17 cases, septic loosening in one case, periprosthetic fracture in another case, osteolysis and polyethylene wear in two cases and, finally, recurrent dislocations in the last one. In the remaining patients, mean HHS increased from a preoperative value of 50,15 to a postoperative value of 92,69. The mean polyethylene wear was 1,25 mm (min. 0,08, max. 3,9 mm), with a mean annual wear of 0,17 mm. The mean acetabular migration on the two axis was 1,6 mm and 1,8 mm. Peri-acetabular osteolysis were recorded in 89% of the implants (163 cases). The cumulative survivorship (revision as endpoint) at the time was 88,9%.</p> <p>Conclusion</p> <p>Our study confirms the bad behavior of this type of cup probably related to the design, to the method of HA fixation. The observations carried out on the revised cup confirm these hypotheses but did not clarify if the third body wear could be a further problem. Another interesting aspect is the high incidence of osteolysis, which are often asymptomatic becoming a problem for the surgeon as the patient refuses the possibility of a revision.</p

    Pregnancy-related pelvic girdle pain: an update

    Get PDF
    A large number of scientists from a wide range of medical and surgical disciplines have reported on the existence and characteristics of the clinical syndrome of pelvic girdle pain during or after pregnancy. This syndrome refers to a musculoskeletal type of persistent pain localised at the anterior and/or posterior aspect of the pelvic ring. The pain may radiate across the hip joint and the thigh bones. The symptoms may begin either during the first trimester of pregnancy, at labour or even during the postpartum period. The physiological processes characterising this clinical entity remain obscure. In this review, the definition and epidemiology, as well as a proposed diagnostic algorithm and treatment options, are presented. Ongoing research is desirable to establish clear management strategies that are based on the pathophysiologic mechanisms responsible for the escalation of the syndrome's symptoms to a fraction of the population of pregnant women

    Eleven-year results of the ABG I hip replacement

    No full text
    The ABG I uncemented hip prosthesis has demonstrated an unacceptably high failure rate and poor functional outcome in our patients. A prospective review was performed of 38 implanted hips with a mean follow-up of 11 years. Sixteen hips have been revised or are candidates for revision hip surgery, due principally to polyethylene wear and pelvic osteolysis. Contributing factors were likely to be the poor wear characteristics of the polyethylene liners—which were gamma irradiated in air—and the thinness of these liners. Of further concern were the poor pain and function scores in the remaining hips [mean Harris Hip Scores (HHS) of 70 (SD 22) at 11-year follow-up]. The influence of the surgical learning curve on clinical results may have been an important factor. Undersized femoral components were associated with poor pain and functional scores in five patients with HHS < 60. All ABG I hip implants must remain under careful, long-term review

    Hip arthroplasty in patients younger than 30 years: excellent ten to 16-year follow-up results with a HA-coated stem

    No full text
    It is well accepted that youth and high activity levels are among the factors that increase the risk of mechanical failure of total hip prostheses. However, there are few reports of long-term results in very young patients. In this study, we evaluated the results of total 49 hip replacements (THRs) using an uncemented total hip prosthesis in 44 patients (28 females) who were 30 years or younger (range: 15–30 years). The diagnosis was ostearthritis due to congenital dislocations in 28 patients, with the remaining patients having diagnoses of sequelae of fracture, infection, Calve-Legg-Perthes disease, avascular necrosis, chondrodystrophia and epiphyseal dysplasia. In all cases we used an uncemented straight stem fully coated with hydroxyapatite (HA). In 36 cases we used a hemispherical cup inserted with press fit, and in seven cases we used a hemispherical screw cup. The patients were evaluated ten to 16 years (mean: 13 years) after the operation by radiographic and clinical examinations, including the Harris Hip, WOMAC and EuroQol-5D scores. In a sub-group of nine patients with a unilateral prosthesis, the muscle strength of the quadriceps and hamstrings was tested using a Cybex 6000. None of the stems were revised at the follow-up examination, and all were classified as well integrated, with no signs of radiological loosening. Twenty-four hips had revision of the acetabular component due to mechanical failure. The Harris Hip score was, on average, 88 (range: 62–100), the WOMAC score 80 (range: 37–100) and the EuroQol score 0.68 (range: −0.14–1). Isokinetic muscle strength testing showed that seven of the nine tested patients were weaker on the operated side. In conclusion, we found mechanical failures at the acetabular side, but excellent results with a fully HA-coated femoral stem, with no revisions after ten to 16 years

    Promising mid-term results of total hip arthroplasties using an uncemented lateral-flare hip prosthesis: a clinical and radiographic study

    No full text
    The clinical results after total hip replacements using noncemented stems have shown considerable variability over the years; the design and characteristics of the implant seemed to play a role in explaining this fact. The purpose of this paper is to report the clinical, radiographic and densitometry results of total hip arthroplasties using a stem designed for noncemented implantation and to engage and load the femur proximally. Fifty-eight consecutive patients (62 hips) followed for an average of 4.3 years (range 36–70 months) were clinically and radiographically followed up at three weeks, three months, six months, one year, and yearly thereafter. The average pre-operative Harris hip score was 49 increasing to 98 at the latest follow-up. There were no cases of aseptic or septic loosening. The average subsidence at three years was 0.45 mm (SD ± 0.36 mm). Radiographically all hips were classified as stable, and evident changes compatible with new bone apposition were observed in 64% of the cases. The extended proximal geometry of the device seems to favour initial and secondary stability as reflected by the low subsidence values over time. The maintenance of periprosthetic bone stock and the absence of stress shielding can be explained by the predominantly proximal loading pattern of the stem
    corecore