    The Association of Central Corneal Thickness and Intraocular Pressure Measures by Non-Contact Tonometry and Goldmann Applanation Tonometry among Glaucoma Patients.

    Get PDF
    BACKGROUND፡ The aim of this study was to determine whether Central Corneal Thickness (CCT) is associated with intraocular pressure measurement (IOP) with a Non-contact tonometer and the Goldmann applanation tonometer in glaucoma patients.MATERIALS AND METHODS: The study involved two hundred and thirty-two eyes of clinically diagnosed glaucoma patients receiving care at a referral facility. IOP measurements were obtained using both the Non-Contact Tonometer (NCT) and Goldmann Applanation Tonometer (GAT). The association between age, ethnicity, and CCT, as well as CCT on the measures of NCT and GAT, were analyzed.RESULTS: There were 64(55.2%) males and 52 (44.8%) females and their ages ranged from 18 to 85 years (mean age = 47.72; SD ±15.75 years). There was a weak positive correlation between the CCT and NCT findings in the right eye (r = 0.19, n = 116, p < 0.05) and in the left eye (r = 0.25, n = 116, p < 0.05). However, there was no correlation between CCT and GAT measurements. Age had a significant correlation with CCT while gender and ethnicity had no significant correlation.CONCLUSION: The study found a significant association between CCT and NCT. However, there was no significant association between CCT and GAT. CCT had an association with age but independent of gender and ethnicity since there was no significant relationship between these variables

    Comparative assessment of the Goldmann applanation and noncontact tonometers in intraocular pressure measurements in a sample of glaucoma patients in the Cape Coast Metropolis, Ghana

    No full text
    Objectives: The objective of the study was to explore the usage of the Goldmann applanation tonometry and noncontact tonometry interchangeably in the measurement of intraocular pressure (IOP) in glaucoma patients. Materials and Methods: The study involved 441 clinically diagnosed glaucoma patients receiving care at a referral facility. IOP measurements were obtained using both the Noncontact tonometer and Goldmann applanation tonometer The repeatability of the measures was analyzed by comparing the repeated measures of the devices using paired t‑test and calculating the correlation coefficient. A Bland–Altman analysis was used to determine the limits of agreement between the two procedures. Results: There were 271 (61.5%) males and 170 (38.5%) females and their age ranged from 18 to 73 years (mean age = 49.37; standard deviation ± 14.81 years). The findings of the study showed significantly lower readings (P < 0.001) of the GAT (right eye = 17.40 ± 7.48 mmHg; left eye = 16.80 ± 7.49 mmHg) compared to the NCT (right eye = 20.15 ± 8.30 mmHg; left eye = 19.74 ± 8.31 mmHg). There was a strong positive correlation between the GAT and NCT findings in the right eye (r = 0.871, n = 441, P < 0.001) and in the left eye (r = 0.887, n = 441, P < 0.001). There was a wide limit of agreement between NCT and GAT measurements. Conclusion: There was statistically significant higher measures obtained with NCT than the GAT but did not exceed the allowable inter‑device difference. There was a strong positive correlation between GAT and NCT measurements. However, it is strongly recommended that these devices are not used interchangeably in the monitoring of IOP in glaucoma due to the wide range of limits of agreement
    corecore