Feasibility of In Vivo Metal Artifact Reduction in Contrast-Enhanced Dedicated Spiral Breast Computed Tomography
Background: Radiopaque breast markers cause artifacts in dedicated spiral breast-computed tomography (SBCT). This study investigates the extent of artifacts in different marker types and the feasibility of reducing artifacts through a metal artifact reduction (MAR) algorithm. Methods: The pilot study included 18 women who underwent contrast-enhanced SBCT. In total, 20 markers of 4 different types were analyzed for artifacts. The extent of artifacts with and without MAR was measured via the consensus of two readers. Image noise was quantitatively evaluated, and the effect of MAR on the detectability of breast lesions was evaluated on a 3-point Likert scale. Results: Breast markers caused significant artifacts that impaired image quality and the detectability of lesions. MAR decreased artifact size in all analyzed cases, even in cases with multiple markers in a single slice. The median length of in-plain artifacts significantly decreased from 31 mm (range 11–51 mm) in uncorrected to 2 mm (range 1–5 mm) in corrected images (p ≤ 0.05). Artifact size was dependent on marker size. Image noise in slices affected by artifacts was significantly lower in corrected (13.6 ± 2.2 HU) than in uncorrected images (19.2 ± 6.8 HU, p ≤ 0.05). MAR improved the detectability of lesions affected by artifacts in 5 out of 11 cases. Conclusion: MAR is feasible in SBCT and improves the image quality and detectability of lesions.</p
Breast Tumor Analysis Using Shifted-Excitation Raman Difference Spectroscopy (SERDS)
We used a shifted-excitation Raman difference spectroscopy method for the ex vivo classification of resected and formalin-fixed
breast tissue samples as normal (healthy) tissue, fibroadenoma, or invasive carcinoma. We analyzed 8 tissue samples containing
invasive carcinoma that were surrounded by normal tissue and 3 tissue samples with fibroadenoma only. We made various measurement sites on various tissue samples, in total 240 measurements for each type of tissue. Although the acquired raw spectra
contain enough information to clearly differentiate between normal and tumor (fibroadenoma and invasive carcinoma) tissue, the
differentiation between fibroadenoma and invasive carcinoma was possible only after the shifted-excitation Raman difference
spectroscopy isolation of pure Raman spectra from the heavily fluorescence interfered raw spectra. We used 784 and 785 nm as
excitation wavelengths for the shifted-excitation Raman difference spectroscopy method. The differences in the obtained pure
Raman spectra are assigned to the different chemical compositions of normal breast tissue, fibroadenoma, and invasive breast
carcinoma. Principal component analysis and linear discriminant analysis showed excellent classification results in the Raman shift
range between 1000 and 1800 cm1
. Invasive breast carcinoma was identified with 99.15% sensitivity, and the absence of invasive
carcinoma was identified with 90.40% specificity. Tumor tissue in tumor-containing tissue was identified with 100% sensitivity, and
the absence of tumor in no-tumor containing tissue was identified with 100% specificity. As gold standard for the determination of
the sensitivity and the specificity, we considered the conventional histopathological classification. In summary, shifted-excitation
Raman difference spectroscopy could be potentially very useful to support histopathological diagnosis in breast patholog
Improving the Quality of Care for Cancer Patients through Oncological Second Opinions in a Comprehensive Cancer Center: Feasibility of Patient-Initiated Second Opinions through a Health-Insurance Service Point
Background: To improve the quality and cost-effectiveness of care, cancer patients can obtain a second medical opinion on their treatment. Validation of the diagnostic procedure (e.g., imaging), diagnosis, and treatment recommendation allows oncological therapy to be applied in a more targeted way, optimizing interdisciplinary care. This study describes patients who received second opinions at the Comprehensive Cancer Center for Erlangen–Nuremberg metropolitan area in Germany over a 6-year period, as well as the amount of time spent on second-opinion counseling. Methods: This prospective, descriptive, single-center observational study included 584 male and female cancer patients undergoing gynecological, urologic, or general surgery who sought a second medical opinion. The extent to which the first opinion complied with standard guidelines was assessed solely descriptively. Results: The first opinion was in accordance with the guidelines and complete in 54.5% of the patients, and guideline compliant but incomplete in 13.2%. The median time taken to form a second opinion was 225 min, and the cancer information service was contacted by patients an average of eight times. Conclusions: The initial opinion was guideline compliant and complete in every second case. Without a second opinion, the remaining patients would have been denied a guideline-compliant treatment recommendation. Obtaining a second opinion gives patients an opportunity to receive a guideline-compliant treatment recommendation and enables them to benefit from newer, individualized therapeutic approaches in clinical trials. Establishing patient-initiated second opinions via central contact points appears to be a feasible option for improving guideline compliance.</jats:p
Discordance between Primary Breast Cancer and Ipsilateral Breast Cancer Tumor Recurrence as a Function of Distance
Background: Risk factors for ipsilateral breast cancer tumor recurrence (IBTR) are well established and include grading, nodal status, and receptor status. Little is known about the influence of the local distance between the primary tumor and recurrences on changes in tumor characteristics and prognosis. Methods: In a retrospective setting, we analyzed primary breast cancers and their recurrences. Localizations of primary and recurrent breast cancer were recorded to calculate the relative distance in pixels. Analysis was performed regarding tumor characteristics, relative distance between both, and their impact on breast cancer prognosis. Results: In a cohort of 142 patients with ipsilateral recurrence, no statistically significant difference could be shown in the change in tumor characteristics depending on distance. Progesterone receptor (PR) and estrogene receptor (ER) status changed in 22.7% and 14.9% of cases, respectively. human epidermal growth factor receptor 2 (ERBB2, HER2) status changed in 18.3% of cases. Survival was in accordance with the literature, with luminal-A-like tumors as best and triple negative breast cancers (TNBC) as worst prognosis. With a threshold of 162 pixels, the survival was significantly better in the group with shorter distance. Conclusion: Change in tumor characteristics from primary breast cancer to recurrence occurs more often in PR than ER. In contrast to other work, in this dataset, recurrences with a larger distance to the primary tumor had a worse prognosis in univariate analysis. A Cox model might indicate the possibility that this influence is independent of other risk factors
Detection of Microcalcifications in Spiral Breast Computed Tomography with Photon-Counting Detector Is Feasible: A Specimen Study
The primary objective of the study was to compare a spiral breast computed tomography system (SBCT) to digital breast tomosynthesis (DBT) for the detection of microcalcifications (MCs) in breast specimens. The secondary objective was to compare various reconstruction modes in SBCT. In total, 54 breast biopsy specimens were examined with mammography as a standard reference, with DBT, and with a dedicated SBCT containing a photon-counting detector. Three different reconstruction modes were applied for SBCT datasets (Recon1 = voxel size (0.15 mm)3, smooth kernel; Recon2 = voxel size (0.05 mm)3, smooth kernel; Recon3 = voxel size (0.05 mm)3, sharp kernel). Sensitivity and specificity of DBT and SBCT for the detection of suspicious MCs were analyzed, and the McNemar test was used for comparisons. Diagnostic confidence of the two readers (Likert Scale 1 = not confident; 5 = completely confident) was analyzed with ANOVA. Regarding detection of MCs, reader 1 had a higher sensitivity for DBT (94.3%) and Recon2 (94.9%) compared to Recon1 (88.5%; p 0.05). The diagnostic confidence of reader 1 was better with SBCT than with DBT (DBT 4.48 ± 0.88, Recon1 4.77 ± 0.66, Recon2 4.89 ± 0.44, and Recon3 4.75 ± 0.72; DBT vs. Recon1/2/3: p < 0.05), while reader 2 found no differences. Sensitivity and specificity for the detection of MCs in breast specimens is equal for DBT and SBCT when a small voxel size of (0.05 mm)3 is used with an equal or better diagnostic confidence for SBCT compared to DBT
Image quality assessment using deep learning in high b-value diffusion-weighted breast MRI
AbstractThe objective of this IRB approved retrospective study was to apply deep learning to identify magnetic resonance imaging (MRI) artifacts on maximum intensity projections (MIP) of the breast, which were derived from diffusion weighted imaging (DWI) protocols. The dataset consisted of 1309 clinically indicated breast MRI examinations of 1158 individuals (median age [IQR]: 50 years [16.75 years]) acquired between March 2017 and June 2020, in which a DWI sequence with a high b-value equal to 1500 s/mm2 was acquired. From these, 2D MIP images were computed and the left and right breast were cropped out as regions of interest (ROI). The presence of MRI image artifacts on the ROIs was rated by three independent observers. Artifact prevalence in the dataset was 37% (961 out of 2618 images). A DenseNet was trained with a fivefold cross-validation to identify artifacts on these images. In an independent holdout test dataset (n = 350 images) artifacts were detected by the neural network with an area under the precision-recall curve of 0.921 and a positive predictive value of 0.981. Our results show that a deep learning algorithm is capable to identify MRI artifacts in breast DWI-derived MIPs, which could help to improve quality assurance approaches for DWI sequences of breast examinations in the future.</jats:p
Evaluation of automated techniques for extraction of circulating cell-free DNA for implementation in standardized high-throughput workflows
AbstractAnalysis of circulating cell-free DNA (ccfDNA) is a suitable tool for detecting somatic mutations for the purpose of making decisions on treatment, monitoring treatment response, and predicting survival. High-throughput techniques for ccfDNA extraction are essential to implementing ccfDNA testing in the clinical setting. We set out to compare two automated techniques with regard to hands-on time, ccfDNA output and integrity, and circulating mitochondrial DNA (mtDNA). CcfDNA was isolated using the EZ1&2 ccfDNA field test kit (EZ2 kit, QIAGEN) and the Maxwell RSC ccfDNA plasma kit (Maxwell kit, Promega). DNA was extracted from plasma of 30 breast cancer patients enrolled in the iMODE-B (#325_19B; 12.10.2020) study. Real-time PCR, fluorescence-based detection and automated electrophoresis were used to assess ccfDNA concentrations. The ccfDNA yield was significantly higher when extracted with the EZ2 kit. The EZ2 kit enabled the isolation of a higher proportion of short fragments and a lower proportion of long fragments, resulting in lower DNA integrity. Significantly lower mtDNA quantities were detected in the Maxwell eluate than in the EZ2 eluate. Thus, decisions on which extraction method to use should proceed on the basis of the required input for downstream applications, the anticipated fragment size and minimum hands-on time.</jats:p
Initial clinical results with a fusion prototype for mammography and three-dimensional ultrasound with a standard mammography system and a standard ultrasound probe
Background
Combinations
*Equal contributors.
of different imaging techniques in fusion devices appear to be associated with improvements in diagnostic assessment.
Purpose
The aim of this study was to test the feasibility of using an automated standard three-dimensional (3D) ultrasound (US) device fused with standard mammography for the first time in breast cancer patients.
Material and Methods
Digital mammograms and 3D automated US images were obtained in 23 patients with highly suspicious breast lesions. A recently developed fusion machine consisting of an ABVS 3D US transducer from an Acuson S2000 machine and a conventional Mammomat Inspiration device (both Siemens Healthcare GmbH, Erlangen, Germany) were used for the purpose. The feasibility of the examinations, imaging coverage, and patients' experience of the procedure were examined.
Results
In 15 out of 19 patients, the region of interest (ROI) with the tumor marked in the mammogram was visible on US. The examination was experienced positively by the patients, with no unexpected pain or injury. The examination was time-saving and well tolerated.
Conclusion
In conclusion, we have shown initial clinical feasibility of an US/radiography fusion prototype with good localization and evaluation of the ROIs. The combined examination was well tolerated. The simultaneous evaluation with mammography and US imaging may be able to improve detection and reduce examiner-related variability
Analysis of motives and patient satisfaction in oncological second opinions provided by a certified university breast and gynecological cancer center
Abstract
Purpose
Second opinions in oncology are becoming increasingly important in an era of more complex treatments and a growing demand for information by patients. Therefore, we analyzed their effects and influencing factors like patients’ motives, subjective extent of information and satisfaction with communications.
Methods
This prospective study evaluated second opinions for patients with breast cancer or gynecological malignancy. The patients received a questionnaire before and two months after, which inquired expectations, reasons, and satisfaction with the second opinion and the attending physicians.
Results
A total of 164 patients were included and the majority had breast cancer (75.0%). Receiving the second opinion made 89.7% feel better informed, their need for information decreased (from 75.3% to 39.2%, P < 0.0001), and satisfaction with doctor–patient communications increased (from 61.9 to 91.8%, P = 0.0002). There were various reasons for requesting a second opinion, e.g., the extremely stressful situation of a cancer diagnosis, hope for change in the treatment recommendation or dissatisfaction with the initial physicians.
Conclusions
Second opinions can lead to significantly greater patient satisfaction, meeting the need for information and leading to better management of patients in the extremely stressful situation of a cancer diagnosis. Doctor–patient communications play a key role
Assessment of the additional clinical potential of X-ray dark-field imaging for breast cancer in a preclinical setup
Background:
Mammography can identify calcifications up to 50–100 μm in size as a surrogate parameter for breast cancer or ductal carcinoma in situ (DCIS). Microcalcifications measuring <50 µm are also associated with breast cancer or DCIS and are frequently not detected on mammography, although they can be detected with dark-field imaging. This study examined whether additional breast examination using X-ray dark-field imaging can increase the detection rate of calcifications. Advances in knowledge:
(1) evaluation of additional modality of breast imaging;
(2) specific evaluation of breast calcifications.
Implications for patient care: the addition of X-ray dark-field imaging to conventional mammography could detect additional calcifications.
Methods:
Talbot–Lau X-ray phase–contrast imaging and X-ray dark-field imaging were used to acquire images of breast specimens. The radiation dosage with the technique is comparable with conventional mammography. Three X-ray gratings with periods of 5–10 µm between the X-ray tube and the flat-panel detector provide three different images in a single sequence: the conventional attenuation image, differential phase image, and dark-field image. The images were read by radiologists. Radiological findings were marked and examined pathologically. The results were described in a descriptive manner.
Results:
A total of 81 breast specimens were investigated with the two methods; 199 significant structures were processed pathologically, consisting of 123 benign and 76 malignant lesions (DCIS or invasive breast cancer). X-ray dark-field imaging identified 15 additional histologically confirmed carcinoma lesions that were visible but not declared suspicious on digital mammography alone. Another four malignant lesions that were not visible on mammography were exclusively detected with X-ray dark-field imaging.
Conclusions:
Adding X-ray dark-field imaging to digital mammography increases the detection rate for breast cancer and DCIS associated lesions with micrometer-sized calcifications.
The use of X-ray dark-field imaging may be able to provide more accurate and detailed radiological classification of suspicious breast lesions.
Adding X-ray dark-field imaging to mammography may be able to increase the detection rate and improve preoperative planning in deciding between mastectomy or breast-conserving therapy, particularly in patients with invasive lobular breast cancer