6 research outputs found
Recommended from our members
Analysis of 83mKr prompt scintillation signals in the PIXeY detector
Prompt scintillation signals from 83mKr calibration sources are a useful metric to calibrate the spatial variation of light collection efficiency and electric field magnitude of a two phase liquid-gas xenon time projection chamber. Because 83mKr decays in two steps, there are two prompt scintillation pulses for each calibration event, denoted S1a and S1b. We study the ratio of S1b to S1a signal sizes in the Particle Identification in Xenon at Yale (PIXeY) experiment and its dependence on the time separation between the two signals (Δ t), notably its increase at low Δ t. In PIXeY data, the Δ t dependence of S1b/S1a is observed to exhibit two exponential components: one with a time constant of 0.05 ± 0.02 μ s, which can be attributed to processing effects and pulse overlap and one with a time constant of 10.2 ± 2.2 μs that increases in amplitude with electric drift field, the origin of which is not yet understood
Recommended from our members
Extraction efficiency of drifting electrons in a two-phase xenon time projection chamber
We present a measurement of the extraction efficiency of quasi-free electrons from the liquid into the gas phase in a two-phase xenon time-projection chamber. The measurements span a range of electric fields from 2.4 to 7.1 kV/cm in the liquid xenon, corresponding to 4.5 to 13.1 kV/cm in the gaseous xenon. Extraction efficiency continues to increase at the highest extraction fields, implying that additional charge signal may be attained in two-phase xenon detectors through careful high-voltage engineering of the gate-anode region
Recommended from our members
Calibration of a two-phase xenon time projection chamber with a 37Ar source
We calibrate a two-phase xenon detector at 0.27 keV in the charge channel and at 2.8 keV in both the light and charge channels using a 37Ar source that is directly released into the detector. We map the light and charge yields as a function of electric drift field. For the 2.8 keV peak, we calculate the Thomas-Imel box parameter for recombination and determine its dependence on drift field. For the same peak, we achieve an energy resolution, Eσ/Emean, between 9.8% and 10.8% for 0.1 kV/cm to 2 kV/cm electric drift fields
Recommended from our members
The electric field dependence of single electron emission in the PIXeY two-phase xenon detector
Dual phase xenon detectors are widely used in experimental searches for galactic dark matter particles. The origin of single electron backgrounds following prompt scintillation and proportional scintillation signals in these detectors is not fully understood, although there has been progress in recent years. In this paper, we describe single electron backgrounds in 83mKr calibration events and their correlation with drift and extraction fields, using the Particle Identification in Xenon at Yale (PIXeY) dual-phase xenon time projection chamber. The single electron background induced by the Fowler-Nordheim (FN) effect is measured, and its electric field dependence is quantified. The photoionization of grids and impurities by prompt scintillation and proportional scintillation also contributes to the single electron background
Recommended from our members
The Electric Field Dependence of Single Electron Emission in the PIXeY Two-Phase Xenon Detector
Dual phase xenon detectors are widely used in experimental searches for
galactic darkmatter particles. The origin of single electron backgrounds
following prompt scintillation and proportional scintillation signals in these
detectors is not fully understood, although there has been progress in recent
years. In this paper, we describe single electron backgrounds in
calibration events and their correlation with drift and extraction fields,
using the Particle Identification in Xenon at Yale (PIXeY) dual-phase xenon
time projection chamber. The single electron background induced by the
Fowler-Nordheim (FN) effect is measured, and its electric field dependence is
quantified. The photoionization of grids and impurities by prompt scintillation
and proportional scintillation also contributes to the single electron
background
Recommended from our members
Analysis of 83mKr prompt scintillation signals in the PIXeY detector
Prompt scintillation signals from 83mKr calibration sources are a useful metric to calibrate the spatial variation of light collection efficiency and electric field magnitude of a two phase liquid-gas xenon time projection chamber. Because 83mKr decays in two steps, there are two prompt scintillation pulses for each calibration event, denoted S1a and S1b. We study the ratio of S1b to S1a signal sizes in the Particle Identification in Xenon at Yale (PIXeY) experiment and its dependence on the time separation between the two signals (Δ t), notably its increase at low Δ t. In PIXeY data, the Δ t dependence of S1b/S1a is observed to exhibit two exponential components: one with a time constant of 0.05 ± 0.02 μ s, which can be attributed to processing effects and pulse overlap and one with a time constant of 10.2 ± 2.2 μs that increases in amplitude with electric drift field, the origin of which is not yet understood