41 research outputs found

    Poor reproducibility of compression elastography in the Achilles tendon: same day and consecutive day measurements.

    Get PDF
    OBJECTIVE To determine the reproducibility of compression elastography (CE) when measuring strain data, a measure of stiffness of the human Achilles tendon in vivo, over consecutive measures, consecutive days and when using different foot positions. MATERIALS AND METHODS Eight participants (4 males, 4 females; mean age 25.5 ± 2.51 years, range 21-30 years; height 173.6 ± 11.7 cm, range 156-189 cm) had five consecutive CE measurements taken on one day and a further five CE measures taken, one per day, at the same time of day, every day for a consecutive 5-day period. These 80 measurements were used to assess both the repeatability and reproducibility of the technique. Means, standard deviations, coefficient of variation (CV), Pearson correlation analysis (R) and intra-class correlation coefficients (ICC) were calculated. RESULTS For CE data, all CVs were above 53%, R values indicated no-to-weak correlations between measures at best (range 0.01-0.25), and ICC values were all classified in the poor category (range 0.00-0.11). CVs for length and diameter measures were acceptably low indicating a high level of reliability. CONCLUSIONS Given the wide variation obtained in the CE results, it was concluded that CE using this specific system has a low level of reproducibility for measuring the stiffness of the human Achilles tendon in vivo over consecutive days, consecutive measures and in different foot positions

    Reproducibility of shear wave elastography measuresof the Achilles tendon.

    Get PDF
    OBJECTIVE To assess the reproducibility of shear wave elastography (SWE) measures in the Achilles tendon (AT) in vivo. MATERIALS AND METHODS Shear wave velocity (SWV) of 14 healthy volunteers [7 males, 7 females; mean age 26.5 ± 3.8 years, mean height 171.6 ± 10.9 cm, mean Victorian Institute of Sports Assessment Achilles questionnaire (VISA-A) score 99.4 ± 1.2] was measured with the foot relaxed and fixed at 90°. Data were collected over five consecutive measures and 5 consecutive days. RESULTS Mean SWV values ranged from 7.91 m/s-9.56 m/s ± 0.27-0.50 m/s. Coefficient of variation (CV), correlations and intra-class correlation coefficient (ICC) scores ranged from 2.9%-6.3%, 0.4-0.7 and 0.54-0.85 respectively. No significant differences were noted for longitudinal or transverse data with respect to protocol or time and no significant differences were noted for foot position in transverse data. Significant differences in SWV values were noted between foot positions for longitudinal scanning (p = <0.05), with a relaxed foot position providing SWV values on average 0.47 m/s faster than a fixed position. Increased reproducibility was obtained with the foot relaxed. ICC between operators was 0.70 for transverse and 0.80 for longitudinal scanning. CONCLUSIONS Reproducible SWE measures were obtained over a 1-h period as well as a period of 5 consecutive days with more reliable measures obtained from a longitudinal plane using a relaxed foot position. SWE also has a high level of agreement between operators making SWE a reproducible technique for quantitatively assessing the mechanical properties of the human AT in vivo
    corecore