    Cultural and molecular evidence of Legionella spp. Colonization in dental unit waterlines: Which is the best method for risk assessment?

    Get PDF
    Legionella spp. are ubiquitous in aquatic habitats and water distribution systems, including dental unit waterlines (DUWLs). The aim of the present study was to determine the prevalence of Legionella in DUWLs and tap water samples using PMA-qPCR and standard culture methods. The total viable counts (TVCs) of aerobic heterotrophic bacteria in the samples were also determined. Legionella spp. were detected and quantified using the modified ISO 11731 culture method. Extracted genomic DNA was analysed using the iQ-Check Quanti Legionella spp. kit, and the TVCs were determined according to the ISO protocol 6222. Legionella spp. were detected in 100% of the samples using the PMA-qPCR method, whereas these bacteria were detected in only 7% of the samples using the culture method. The number of colony forming units (CFUs) of the TVCs in the DUWL and tap water samples differed, with the bacterial load being significantly lower in the tap water samples (p-value = 0). The counts obtained were within the Italian standard range established for potable water in only 5% of the DUWL water samples and in 77% of the tap water samples. Our results show that the level of Legionella spp. contamination determined using the culture method does not reflect the true scale of the problem, and consequently we recommend testing for the presence of aerobic heterotrophic bacteria based on the assumption that Legionella spp. are components of biofilms

    The role of chemical products at low doses in preventing the proliferation of bacteria in dental unit waterlines: the ICX® experience.

    Get PDF
    Abstract In this study we evaluated (1) the efficacy of a protocol that combines hydrogen peroxide (shock treatment) and ICX® tablets (continuous treatment) for the control of microbial contamination in dental unit water lines, and (2) the in vitro antimicrobial activity of ICX® tablets on collection and wild strains isolated from dental chair output waters. To assess the treatment effectiveness, the microbial load in the output water samples of three dental chairs were investigated: one control chair received only shock treatment. In vitro bactericidal activity was tested against Staphylococcus aureus and Pseudomonas aeruginosa. Data obtained from samples collected from chairs treated with ICX® and shock treatment and data from the control chair did not differ significantly on the basis of microbial load. In the in vitro study, the product was unable to kill Gram-negative bacteria. These results show that the continuous introduction of ICX® was not effective in maintaining low counts of the heterotrophic bacteria in the output water of dental devices, and shock treatment may be needed more frequently than monthly

    Experimental Study to Develop a Method for Improving Sample Collection to Monitor Laryngoscopes after Reprocessing

    Get PDF
    Background/Aims The microbiological surveillance of endoscopes and automated flexible endoscope reprocessing have been proven to be two of the most difficult and controversial areas of infection control in endoscopy. The purpose of this study was to standardize a sampling method for assessing the effectiveness of standard reprocessing operating procedures for flexible fiberoptic laryngoscopes (FFLs). Methods First, the sampling devices were directly inoculated with Bacillus atrophaeus spores; second, tissue non tissue (TNT) wipes were tested on artificially contaminated surfaces and on FFLs. Results Comparison of the sponges, cellulose, and TNT wipes indicated that the TNT wipes were more effective in releasing spores (93%) than the sponges (49%) and cellulose wipes (52%). The developed protocol provides a high efficiency for both collection and extraction from the stainless steel surface (87% of the spores were removed and released) and from the FFL (85% of the spores were removed and released), with relatively low standard deviations for recovery efficiency, particularly for the analysis of the FFL. Conclusions TNT wipes are more efficient for sampling surface areas, thereby aiding in the accuracy and reproducibility of environmental surveillance

    Seroprevalence of infection-induced SARS-CoV-2 antibodies among health care users of Northern Italy: results from two serosurveys (October-November 2019 and September-October 2021)

    Get PDF
    OBJECTIVES: The objective was to estimate the seroprevalence of SARS-CoV-2 in autumn 2019 (before case zero was identified in Italy) and 2021 among residual sera samples from health care users in the Piedmont region of northwestern Italy. METHODS: Two serosurveys were conducted. Using a semiquantitative method, samples were tested for the presence of immunoglobulin G (IgG) antibodies against the S1 domain of the spike protein. Samples with positive test results from the 2019 survey were independently retested using a multiplex panel to detect IgG antibodies against the receptor binding domain, S1 and S2 domains, and nucleocapsid. Samples with positive test results from the 2021 survey underwent repeat testing with enzyme-linked immunosorbent assay to detect anti-nucleocapsid IgG antibodies. Prevalence rates according to gender and age groups, together with their respective 95% confidence intervals (CIs), were calculated. RESULTS: Overall, the proportion of samples with positive test results was 2/353 in 2019 and 22/363 in 2021, with an estimated seroprevalence of 0.27% (95% CI 0-1.86) and 6.21% (95% CI 3.9-9.31) in 2019 and 2021 respectively. CONCLUSION: Results of this study support the hypothesis that the virus was circulating in Italy as early as autumn 2019. The role of these early cases in broader transmission dynamics remains to be determined

    Cultural and Molecular Evidence of Legionella spp. Colonization in Dental Unit Waterlines: Which Is the Best Method for Risk Assessment?

    No full text
    Legionella spp. are ubiquitous in aquatic habitats and water distribution systems, including dental unit waterlines (DUWLs). The aim of the present study was to determine the prevalence of Legionella in DUWLs and tap water samples using PMA-qPCR and standard culture methods. The total viable counts (TVCs) of aerobic heterotrophic bacteria in the samples were also determined. Legionella spp. were detected and quantified using the modified ISO 11731 culture method. Extracted genomic DNA was analysed using the iQ-Check Quanti Legionella spp. kit, and the TVCs were determined according to the ISO protocol 6222. Legionella spp. were detected in 100% of the samples using the PMA-qPCR method, whereas these bacteria were detected in only 7% of the samples using the culture method. The number of colony forming units (CFUs) of the TVCs in the DUWL and tap water samples differed, with the bacterial load being significantly lower in the tap water samples (p-value = 0). The counts obtained were within the Italian standard range established for potable water in only 5% of the DUWL water samples and in 77% of the tap water samples. Our results show that the level of Legionella spp. contamination determined using the culture method does not reflect the true scale of the problem, and consequently we recommend testing for the presence of aerobic heterotrophic bacteria based on the assumption that Legionella spp. are components of biofilms

    Persistence of Legionella in Routinely Disinfected Heater-Cooler Units and Heater Units assessed by Propidium Monoazide qPCR

    No full text
    Background: Evidence to date indicates that heater–cooler units (HCUs) and heater units (HUs) can generate potentially infectious aerosols containing a range of opportunistic pathogens such as Mycobacterium chimaera, other non-tuberculous mycobacterial (NTM) species, Pseudomonas aeruginosa and Legionella spp. Our purpose was to determine the extent of Legionella contamination and total viable count (TVC) in HCUs and HUs and to analyze the relationship by water system design of devices of two different brands (LivaNova vs. Maquet). Methods: Legionella spp. were detected and quantified by our optimized PMA-qPCR protocol; TVCs were assessed according to ISO protocol 6222. Analyses were performed in the first sampling round and after six months of surveillance. Results: Overall, Legionella spp. was detected in 65.7% of devices. In the second sampling round, Legionella positivity rates were significantly lower in water samples from the Maquet devices compared to the LivaNova ones (27.3% vs. 61.5%). LivaNova HCUs also yielded more Legionella, and aquatic bacteria counts than Maquet in both first and second-round samples. Conclusions: We recommend that all surgical patients and staff exposed to aerosols from thermoregulatory devices should be followed up for Legionella infection and that microbiological surveillance on such devices should be conducted regularly as precautionary principle