    Semicontinuous intra-abdominal pressure measurement using an intragastric Compliance catheter

    Get PDF
    OBJECTIVE: To compare intra-abdominal pressure (IAP) measurements obtained from an intragastric Compliance catheter with the pressure measured directly in the abdominal cavity. DESIGN AND SETTING: Prospective cohort study in an operating room of the Ghent University Hospital PATIENTS: Seven patients undergoing elective laparoscopic cholecystectomy. INTERVENTIONS: IAP was obtained from both an intragastric catheter and directly from the peritoneal cavity at 1-minute intervals in patients undergoing elective cholecystectomy and compared using Bland-Altman analysis. MEASUREMENTS AND RESULTS: In 156 paired measurements obtained from 7 patients the mean difference between IAPgastric and IAPref was 0.12+/-0.70 mmHg (95% CI 0.01-0.23). CONCLUSIONS: IAP measured using an intragastric Compliance catheter reliably reflects the reference IAP in patients undergoing laparoscopic cholecystectomy