9 research outputs found

    Different glucose analyzers report different glucose concentration values in term newborns

    Get PDF
    Background: The American Academy of Pediatrics and Pediatric Endocrine Society neonatal hypoglycemia guidelines based their glucose concentration treatment thresholds on studies that predominantly used Beckman and Yellow Springs Glucose Oxidase Analyzers. Currently, a majority (76%) of U.S. hospital laboratories utilizing glucose oxidase methodology use Vitros® Glucose Analyzers. However, a bias of ~+5% between glucose concentrations from Beckman vs. Vitros Glucose Analyzers has been reported; this could have a clinically significant effect when using published guideline treatment thresholds. Methods: To determine if there is similar instrument bias between Beckman and Vitros Analyzers in reported glucose concentrations from term newborns, we compared plasma glucose concentrations measured within the first 3 h after birth by Beckman vs. Vitros Analyzers in a total of 1,987 newborns (Beckman n = 904, Vitros n = 1,083). Data were fit using nonlinear cubic spline models between collection time and glucose concentration. Results: The non-linear patterns of initial glucose concentrations (during the first 3 h after birth) as measured by Beckman and Vitros Analyzers paralleled each other with no overlap of the fit spline curve 95% confidence intervals, with an approximate +5 mg/dL constant bias. Additionally, in method comparison studies performed in the Chemistry Laboratory on adult samples, there was a +4.2-7.4 mg/dL measured glucose bias for the Beckman vs. Vitros Analyzer. Conclusion: Glucose concentrations from term, appropriate size for gestational age newborns were about 5 mg/dL higher when measured by Beckman vs. Vitros Analyzers. Perhaps, concentrations of 45 mg/dL reported from Beckman Analyzers may be equivalent to 40 mg/dL from Vitros Analyzers. When managing neonatal hypoglycemia, it is important to know which analyzer was used and whether adjusting for potential instrument bias is necessary when following published guidelines. Keywords: glucose; glucose analyzer; guideline; neonatal hypoglycemia; newborn

    P-tau/aβ42 and aβ42/40 ratios in csf are equally predictive of amyloid pet status

    No full text
    Introduction: Measurement of amyloid beta (Aβ40 and Aβ42) and tau (phosphorylated tau [p-tau] and total tau [t-tau]) in cerebrospinal fluid (CSF) can be utilized to differentiate clinical and preclinical Alzheimer’s disease dementia (AD) from other neurodegenerative processes. Methods: CSF biomarkers were measured in 150 participants from the Mayo Clinic Study of Aging and the Alzheimer’s Disease Research Center. P-tau/Aβ42 (Roche Elecsys, Fujirebio LUMIPULSE) and Aβ42/40 (Fujirebio LUMIPULSE) ratios were compared to one another and to amyloid positron emission tomography (PET) classification. Results: Strong correlation was observed between LUMIPULSE p-tau/Aβ42 and Aβ42/40, as well as Elecsys and LUMIPULSE p-tau/Aβ42 and Aβ42/40 (Spearman’s ρ = –0.827, –0.858, and 0.960, respectively). Concordance between LUMIPULSE p-tau/Aβ42 and Aβ42/40 was 96% and between Elecsys p-tau/Aβ42 and both LUMIPULSE ratios was 97%. All ratios had > 94% overall, positive, and negative percent agreement with amyloid PET classification. Discussion: These data suggest that p-tau/Aβ42 and Aβ42/40 ratios provide similar clinical information in the assessment of amyloid pathology

    Abstracts of Presentations at the Association of Clinical Scientists 139

    No full text
    corecore