    An assessment of biomedical CoCrMo alloy fabricated by direct metal laser sintering technique for implant applications

    No full text
    CoCrMo alloys have been used for several decades in implantable devices due to their favourable mechanical properties, low wear rate in addition to good biocompatibility and high corrosion resistance. These alloys are conventionally produced via casting and/or forging route, however additive manufacturing techniques being recently employed in their fabrication. In this work, CoCrMo samples were produced by direct metal laser sintering additive manufacturing process. The microstructure and surface composition were examined employing scanning electron microscopy, X-ray diffraction and X-ray photoelectron spectroscopy (XPS). The corrosion resistance was measured in 0.14 M sodium chloride solution and in phosphate buffered solution (PBS) both with and without addition of albumin at pH 7.4 and 37 degrees C. For this, potentiodynamic tests in addition to electrochemical impedance spectroscopy were employed. The studied CoCrMo alloy exhibits a good corrosion resistance in solutions tested being the highest in PBS solution without albumin addition. The XPS analysis showed that the passive film composition and its thickness are not modified by the adsorbed layer. Microstructural analysis revealed occurrence of strain-induced martensitic transformation107COORDENAÇÃO DE APERFEIÇOAMENTO DE PESSOAL DE NÍVEL SUPERIOR - CAPESFUNDAÇÃO DE AMPARO À PESQUISA DO ESTADO DE SÃO PAULO - FAPESP50/20112016/09350-2; 2007/54829-