    NaI(Tl) Scintillator's Response Functions for Point-like and Distributed Gamma-ray Sources

    Get PDF
    The response functions of a NaI(Tl) detectors have been estimated using Monte Carlo methods. Response functions were calculated for monoenergetic photon sources (0.05 to 3 MeV). Responses were calculated for point-like sources and for sources distributed in Portland cement cylinders. Calculated responses were used to estimate the detector efficiency for point-like and distributed sources. Samples of cylindrical Portland cement were prepared and exposed to the photoneutron field produced by a 15 MV linac used for radiotherapy. Short half-life radioisotopes were induced and the activity was determined by measuring the pulse-height spectra with a NaI(Tl) g-ray spectrometer that was calibrated using point-like sources. Instead of doing corrections due to differences between the geometry, material and solid angle of point-like sources used for calibration, and the Portland cement cylinders, the detection efficiency was determined using the ratio between the efficiencies for the point-like and the distributed sources estimated with the Monte Carlo calculations, and the activity of the induced isotopes in cement was obtained