    Status of Aqua MODIS On-orbit Calibration and Characterization

    Get PDF
    The MODIS Flight Model 1 (FM1) has been in operation for more than two years since its launch onboard the NASA's Earth Observing System (EOS) Aqua spacecraft on May 4, 2002. The MODIS has 36 spectral bands: 20 reflective solar bands (RSB) with center wavelengths from 0.41 to 2.2 micron and 16 thermal emissive bands (TEB) from 3.7 to 14.5 micron. It provides the science community observations (data products) of the Earth's land, oceans, and atmosphere for a board range of applications. Its primary on-orbit calibration and characterization activities are performed using a solar diffuser (SD) and a solar diffuser stability monitor (SDSM) system for the RSB and a blackbody for the TEB. Another on-board calibrator (OBC) known as the spectro-radiometric calibration assembly (SRCA) is used for the instrument's spatial (TEB and RSB) and spectral (RSB only) characterization. We present in this paper the status of Aqua MODIS calibration and characterization during its first two years of on-orbit operation. Discussions will be focused on the calibration activities executed on-orbit in order to maintain and enhance the instrument's performance and the quality of its Level 1B (L1B) data products. We also provide comparisons between Aqua MODIS and Terra MODIS (launched in December, 1999), including their similarity and difference in response trending and optics degradation. Existing data and results show that Aqua MODIS bands 8 (0.412 micron) and 9 (0.443 micron) have much smaller degradation than Terra MODIS bands 8 and 9. The most noticeable feature shown in the RSB trending is that the mirror side differences in Aqua MODIS are extremely small and stable (<0.1%) while the Terra MODIS RSB trending has shown significant mirror side difference and wavelength dependent degradation. The overall stability of the Aqua MODIS TEB is also better than that of the Terra MODIS during their first two years of on-orbit operation

    MODIS Instrument Operation and Calibration Improvements

    Get PDF
    Terra and Aqua MODIS have successfully operated for over 14 and 12 years since their respective launches in 1999 and 2002. The MODIS on-orbit calibration is performed using a set of on-board calibrators, which include a solar diffuser for calibrating the reflective solar bands (RSB) and a blackbody for the thermal emissive bands (TEB). On-orbit changes in the sensor responses as well as key performance parameters are monitored using the measurements of these on-board calibrators. This paper provides an overview of MODIS on-orbit operation and calibration activities, and instrument long-term performance. It presents a brief summary of the calibration enhancements made in the latest MODIS data collection 6 (C6). Future improvements in the MODIS calibration and their potential applications to the S-NPP VIIRS are also discussed

    Comparison of MODIS and VIIRS On-board Blackbody Performance

    Get PDF
    MODIS has 16 thermal emissive bands (TEBs), covering wavelengths from 3.7 to 14.4 microns. MODIS TEBs are calibrated on-orbit by a v-grooved blackbody (BB) on a scan-by-scan basis. The BB temperatures are measured by a set of 12 thennistors. As expected, the BB temperature uncertainty and stability have direct impact on the quality of TEB calibration and, therefore, the quality of the science products derived from TEB observations. Since launch, Terra and Aqua MODIS have successfully operated for more than 12 and 10 years, respectively. Their on-board BB performance has been satisfactory in meeting the TEB calibration requirements. The first VIIRS, launched on-board the Suomi NPP spacecraft on October 28, 2011, has successfully completed its initial Intensive Calibration and Validation (ICV) phase. VIIRS has 7 thermal emissive bands (TEBs), covering wavelengths from 3.7 to 12.4 microns. Designed with strong MODIS heritage, VIIRS uses a similar BB for its TEB calibration. Like MODIS, VIIRS BB is nominally controlled at a pre-determined temperature (set point). Periodically, a BB Warm-Up and Cool-Down (WUCD) operation is performed, during which the BB temperatures vary from instrument ambient (temperature) to 315K. This paper examines NPP VIIRS BB on-orbit performance. It focuses on its BB temperature scan-to-scan variations at nominally controlled temperature as well as during its WUCD operation and their impact on TEB calibration uncertainty. Comparisons of VIIRS (NPP) and MODIS (Terra and Aqua) BB on-orbit performance and lessons learned for future improvements are also presented in this paper

    MODIS On-orbit Calibration Uncertainty Assessment

    Get PDF
    MODIS has 20 reflective solar bands (RSB) and 16 thermal emissive bands (TEB). Compared to its heritage sensors, MODIS was developed with very stringent calibration uncertainty requirements. As a result, MODIS was designed and built with a set of on-board calibrators (OBC), which allow key sensor performance parameters and on-orbit calibration coefficients to be monitored and updated. In terms of its calibration traceability, MODIS RSB calibration is reflectance based using an on-board solar diffuser (SD) and the TEB calibration is radiance based using an on-board blackbody (BB). In addition to on-orbit calibration coefficients derived from its OBC, calibration parameters determined from sensor pre-launch calibration and characterization are used in both the RSB and TEB calibration and retrieval algorithms. This paper provides a brief description of MODIS calibration methodologies and an in-depth analysis of its on-orbit calibration uncertainties. Also discussed in this paper are uncertainty contributions from individual components and differences due to Terra and Aqua MODIS instrument characteristics and on-orbit performance

    Aqua MODIS Thermal Emissive Band On-Orbit Calibration, Characterization, and Performance

    Get PDF
    The NASA's Earth Observing System Aqua Moderate Resolution Imaging Spectroradiometer (MODIS) has continued to operate with satisfactory performance since its launch in May 2002, exceeding its nominal six-year design lifetime. Its continuous Earth observations have been used to generate many science data products for studies of the Earth's, system. MODIS has 36 spectral bands; 20 reflective solar bands and 16 thermal emissive bands (TEBs). All TEB observations are made at 1-km nadir spatial resolution with spectral wavelengths from 3.7 to 14.4 pm. Primary applications of MODIS TEB 'include surface, cloud, and atmospheric temperatures, water vapor, and cloud top altitude. MODIS TEB on-orbit calibration uses a quadratic algorithm with its calibration coefficients derived using an onboard blackbody (BB). This paper will present Aqua MODIS TEB on-orbit calibration, characterization, and performance over its six-year mission. Examples of instrument thermal behavior, BB temperature stability, detector short-term stability, and changes in long-term response (or system gain) will be presented. Comparisons will also be made with Terra MODIS, launched in December 1999. On-orbit results show that Aqua MODIS and its focal plane temperatures have behaved normally. BB temperature has remained extremely stable with typical scan -to-scan variations of less than +/-0.15 mK. Most TEB detectors continue to exceed their specified signal-to-noise ratio requirements, exhibiting excellent short-term stability and calibration accuracy. Excluding a few noisy detectors, either identified prelaunch or occurring postlaunch, on-orbit changes in TEB responses have been less than 0.5% on an annual basis. By comparison; the overall Aqua TEB performance has been better than that of Terra MODIS

    MODIS On-Board Blackbody Function and Performance

    Get PDF
    Two MODIS instruments are currently in orbit, making continuous global observations in visible to long-wave infrared wavelengths. Compared to heritage sensors, MODIS was built with an advanced set of on-board calibrators, providing sensor radiometric, spectral, and spatial calibration and characterization during on-orbit operation. For the thermal emissive bands (TEB) with wavelengths from 3.7 m to 14.4 m, a v-grooved blackbody (BB) is used as the primary calibration source. The BB temperature is accurately measured each scan (1.47s) using a set of 12 temperature sensors traceable to NIST temperature standards. The onboard BB is nominally operated at a fixed temperature, 290K for Terra MODIS and 285K for Aqua MODIS, to compute the TEB linear calibration coefficients. Periodically, its temperature is varied from 270K (instrument ambient) to 315K in order to evaluate and update the nonlinear calibration coefficients. This paper describes MODIS on-board BB functions with emphasis on on-orbit operation and performance. It examines the BB temperature uncertainties under different operational conditions and their impact on TEB calibration and data product quality. The temperature uniformity of the BB is also evaluated using TEB detector responses at different operating temperatures. On-orbit results demonstrate excellent short-term and long-term stability for both the Terra and Aqua MODIS on-board BB. The on-orbit BB temperature uncertainty is estimated to be 10mK for Terra MODIS at 290K and 5mK for Aqua MODIS at 285K, thus meeting the TEB design specifications. In addition, there has been no measurable BB temperature drift over the entire mission of both Terra and Aqua MODIS

    Overview of NASA Earth Observing Systems Terra and Aqua Moderate Resolution Imaging Spectroradiometer Instrument Calibration Algorithms and On-Orbit Performance

    Get PDF
    Since launch, the Terra and Aqua moderate resolution imaging spectroradiometer (MODIS) instruments have successfully operated on-orbit for more than 9 and 6.5 years, respectively. Iv1ODIS, a key instrument for the NASA's Earth Observing System (EOS) missions, was designed to make continuous observations for studies of Earth's land, ocean, and atmospheric properties and to extend existing data records from heritage earth-observing sensors. In addition to frequent global coverage, MODIS observations are made in 36 spectral bands, covering both solar reflective and thermal emissive spectral regions. Nearly 40 data products are routinely generated from MODIS' observations and publicly distributed for a broad range of applications. Both instruments have produced an unprecedented amount of data in support of the science community. As a general reference for understanding sensor operation and calibration, and thus science data quality, we ;provide an overview of the MODIS instruments and their pre-launch calibration and characterization, and describe their on-orbit calibration algorithms and performance. On-orbit results from both Terra and Aqua MODIS radiometric, spectral, and "spatial calibration are discussed. Currently, both instruments, including their on-board calibration devices, are healthy and are expected to continue operation for several }rears to come

    Terra and Aqua MODIS TEB Inter-Comparison Using Himawari-8/AHI as Reference

    Get PDF
    Intercomparison between the two MODIS instruments is very useful for both the instrument calibration and its uncertainty assessment. Terra and Aqua MODIS have almost identical relative spectral response, spatial resolution, and dynamic range for each band, so the site-dependent effect from spectral mismatch for their comparison is negligible. Major challenges in cross-sensor comparison of instruments on different satellites include differences in observation time and view angle over selected pseudoinvariant sites. The simultaneous nadir overpasses (SNO) between the two satellites are mostly applied for comparison and the scene under SNO varies. However, there is a dearth of SNO between the Terra and Aqua. This work focuses on an intercomparison method for MODIS thermal emissive bands using Himawari-8 Advanced Himawari Imager (AHI) as a reference. Eleven thermal emissive bands on MODIS are at least to some degree spectrally matched to the AHI bands. The sites selected for the comparison are an ocean area around the Himawari-8 suborbital point and the Strzelecki Desert located south of the Himawari-8 suborbital point. The time difference between the measurements from AHI and MODIS is &lt;5 min. The comparison is performed using 2017 collection 6.1 L1B data for MODIS. The MODISAHI difference is corrected to remove the view angle dependence. The TerraAqua MODIS difference for the selected TEB is up to 0.6 K with the exception of band 30. Band 30 has the largest difference, which is site dependent, most likely due to a crosstalk effect. Over the ocean, the band 30 difference between the two MODIS instruments is around 1.75 K, while over the desert; the difference is around 0.68 K. The MODIS precision is also compared from the Gaussian regression of the double difference. Terra bands 27 to 30 have significant extra noise due to crosstalk effects on these bands. These TerraAqua comparison results are used for MODIS calibration assessments and are beneficial for future calibration algorithm improvement. The impact of daytime measurements and the scene dependence are also discussed

    An Overview of MODIS On-orbit Operation, Calibration, and Lessons

    Get PDF
    Two nearly identical copies of the Moderate Resolution Imaging Spectroradiometer (MODIS) have successfully operated onboard the Terra and Aqua spacecraft for more than II years and 9 years since their launch in December 1999 and May 2002, respectively. MODIS is a key instrument for the NASA's Earth Observing System (EOS) missions. MODIS observations have produced an unprecedented amount and a broad range of data products and significantly benefited the science and user community. Its follow-on instrument, the Visible/Infrared Imager Radiometer Suite (VIIRS) on-board the NPOESS Preparatory Project (NPP) spacecraft, is currently scheduled for launch in late October, 2011. The NPP serves as a bridge mission between EOS and the Joint Polar Satellite System (JPSS). MODIS collects data in 36 spectral bands, covering spectral regions from visible (VIS) to long-wave infrared (L WIR), and at three different spatial resolutions. Because of its stringent design requirements, MODIS was built with a complete set of onboard calibrators, including a solar diffuser (SO), a solar diffuser stability monitor (SDSM), a blackbody (BB), a spectroradiometric calibration assembly (SRCA), and a space view (SV) port. Except for tbe SRCA, VIlRS carries the same set of onboard calibrators as MODIS. The SD/SDSM system is used together to calibrate tbe reflective solar bands (RSB). The BB is designed for the thermal emissive bands (TEB) calibration. Similar to Terra and Aqua MODIS, VIlRS will also make regular lunar observations to monitor RSB radiometric calibration stability. In this paper, we provide an overview of MODIS on-orbit operation and calibration activities and present issues identified and lessons learned from mission-long instrument operations and implementation of various calibration and characterization strategies. Examples of both Terra and Aqua MODIS instrument on-orbit performance, including their similarities and unique characteristics, are discussed in tbe context of what might be expected from and benefited to tbe NPP VIlRS operation and calibration. It is anticipated that MODIS experience and lessons will also provide valuable information for other earth observing missions/sensors
    • …