Impact of Audiovisual-Assisted Therapeutic Ambience in Radiation Therapy (AVATAR) on Anesthesia Use, Payer Charges, and Treatment Time in Pediatric Patients
Purpose
Pediatric radiation therapy (RT) requires optimal immobilization that often necessitates daily anesthesia. To decrease anesthesia use, we implemented a novel audiovisual-assisted therapeutic ambience in RT (AVATAR) system that projects video onto a radiolucent screen within the child’s line of vision to provide attentional diversion. We investigated its reduction on anesthesia use, payer charges, and treatment time, in addition to its impact on radiation delivery.
Methods and Materials
A 6-year retrospective analysis was performed among children undergoing RT (n = 224) 3 years before and 3 years after the introduction of AVATAR. The frequency of anesthesia use before and after AVATAR implementation, in addition to RT treatment times, were compared. The number of spared anesthesia treatments allowed for a charge to payer analysis. To document the lack of surface dose perturbation by AVATAR, a phantom craniospinal treatment course was delivered both with and without AVATAR. Additionally, an ion chamber course was delivered to document changes to the dose at depth.
Results
More children were able to avoid anesthesia use entirely in the post-AVATAR cohort compared with the pre-AVATAR cohort (73.2% vs 63.4%; P = .03), and fewer required anesthesia for each treatment (18.8% vs 33%; P = .03). AVATAR introduction reduced anesthesia use for all ages studied. Treatment time per session was reduced by 38% using AVATAR compared with anesthesia. There were 326 fewer anesthesia sessions delivered over 3 years after AVATAR was introduced, with an estimated savings of >500,000.Opticallystimulatedluminescentdosimetersrevealedasmallincreaseindoseof0.8ConclusionsAVATARintroductiondecreasedanesthesiauseinchildrenundergoingRT.Morechildrenavoidedanesthesiaentirely,andfewerneededanesthesiaforeverytreatment,resultinginareductionintreatmenttimeandsavingsofnearly550,000 in approximately 3 years, with minimal perturbation of RT dose delivery
Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.