High Performance Liquid Chromatography Assay Method for Simultaneous Quantitation of Formoterol and Budesonide in Symbicort Turbuhaler

Abstract

NoA sensitive and rapid high performance liquid chromatography method has been developed and used for the simultaneous determination of formoterol and budesonide in Symbicort Turbuhaler when assessing the aerodynamic characteristics of the emitted dose using Pharmacopoeial methods. This capability results in both time and cost saving. The mobile phase composition was acetonitrile-5 mM sodium dihydrogen orthophosphate, pH 3 (60: 40% v/v), and was passed at 1.5 ml min-1 through a C18 column with a UV detection (wavelength 214 nm). The method was shown to give good analytical performance in terms of linearity, precision (using phenylpropanolamine as an internal standard), sensitivity and solution stability. The intra-day precision for both formoterol and budesonide were 0.75% and 1.11%, respectively (n = 10). The limit of quantitation for formoterol was 10 ¿g L-1 and for budesonide was 120 ¿g L-1, and the limit of detection were 3 and 30 ¿g L-1, for both formoterol and budesonide, respectively. The method has been applied to determine the content of the emitted dose and the fine particle dose of Symbicort Turbuhaler

Similar works

Full text

thumbnail-image

Bradford Scholars

redirect

This paper was published in Bradford Scholars.

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.