The HERA-B RICH uses a radiation path length of 2.8 m in C_4F_10 gas and a
large 24 square meters spherical mirror for imaging Cherenkov rings. The photon
detector consists of 2240 Hamamatsu multi-anode photomultipliers with about
27000 channels. A 2:1 reducing two-lens telescope in front of each PMT
increases the sensitive area at the expense of increased pixel size, resulting
in a contribution to the resolution which roughly matches that of dispersion.
The counter was completed in January of 1999, and its performance has been
steady and reliable over the years it has been in operation. The design
performance of the RICH was fully reached: the average number of detected
photons in the RICH for a beta=1 particle was found to be 33 with a single hit
resolution of 0.7 mrad and 1 mrad in the fine and coarse granularity regions,
respectively.Comment: 29 pages, 23 figure