Investigation of Non-linear Chirp Coding for Improved Second Harmonic Pulse Compression

Abstract

Non-linear frequency-modulated (NLFM) chirp coding was investigated to improve the pulse compression of the second harmonic chirp signal by reducing the range side lobe level. The problem of spectral overlap between the fundamental component and second harmonic component (SHC) was also investigated. Therefore, two methods were proposed: method I for the non-overlap condition and method II with the pulse inversion technique for the overlap harmonic condition. In both methods, the performance of the NLFM chirp was compared with that of the reference LFM chirp signals. Experiments were performed using a 2.25 MHz transducer mounted coaxially at a distance of 5 cm with a 1 mm hydrophone in a water tank, and the peak negative pressure of 300 kPa was set at the receiver. Both simulations and experimental results revealed that the peak side lobe level (PSL) of the compressed SHC of the NLFM chirp was improved by at least 13 dB in method I and 5 dB in method II when compared with the PSL of LFM chirps. Similarly, the integrated side lobe level (ISL) of the compressed SHC of the NLFM chirp was improved by at least 8 dB when compared with the ISL of LFM chirps. In both methods, the axial main lobe width of the compressed NLFM chirp was comparable to that of the LFM signals. The signal-to-noise ratio of the SHC of NLFM was improved by as much as 0.8 dB, when compared with the SHC of the LFM signal having the same energy level. The results also revealed the robustness of the NLFM chirp under a frequency-dependent attenuation of 0.5 dB/cm·MHz up to a penetration depth of 5 cm and a Doppler shift up to 12 kHz