1325-1331A new calixarene
derivative, i.e., p-tert-butylcalix[4]arene
has been used as an excellent ionophore in the construction of a novel PVC
based Eu(III)-selective electrode. A master membrane with percentage
composition of PVC:DBP:L: NaTPB:: 32:55:4:9 gives the best results. The
membrane electrode exhibits a nernstian response with the slope of
19.7±1.0 mV/decade over a wide concentration range of 9.1 × 10-8-1.0
× 10-1 M and with
detection limit of 8.8 × 10-8 M.
The effects of membrane composition and pH
as well as the influence of the anionic additive on the electrode performance
have been investigated. The potential response of the resulting sensor is not
influenced in the pH range 3.0-9.0.
Moreover, the membrane electrode has a good reproducibility (RSD 3%), very fast
response time (5 s) and ability to detect Eu(III) among other rare earth metal
and common cations. The membrane sensor works satisfactorily up to 40%
non-aqueous medium and can be used over a period of 6 months. The sensor has
been successfully applied for the determination of Eu(III) in analytical and
biological samples (urine and serum) and also acts as an electrode indicator