Selective amperometric sensing of hydrogen peroxide with Nafion/copper particulates chemically modified electrode

Abstract

940-945Nafion/copper particulates chemically modified glassy carbon electrode has been prepared by in situ electrochemical deposition. The electrode is found to to be successful for selective amperometric sensing of hydrogen peroxide at -0.2 V vs Ag/AgCl at physiological pH without interference from uric acid, ascorbic acid, catechol, cysteine, nitrite and nitrate. Two linear regions, 20-200 µM and 220-400 µM, for H2O2 detection with the slope and regression coefficient values of 33.2 nA/µM and 25.61 nA/µM and 0.9975 and 0.9999 respectively are obtained with 1.63 µM detection limit (S/N = 3). Application to real samples analysis is demonstrated for milk, urine and tea samples with appreciable recovery values

    Similar works

    Full text

    thumbnail-image

    Available Versions

    Last time updated on 11/04/2020