Spectra from the Short Wavelength Spectrometer (SWS) on ISO exhibit artifacts
at 4.5 and 8 um. These artifacts appear in spectra from a recent data release,
OLP 10.0, as spurious broad emission features in the spectra of stars earlier
than ~F0, such as alpha CMa. Comparison of absolutely calibrated spectra of
standard stars to corresponding spectra from the SWS reveals that these
artifacts result from an underestimation of the strength of the CO and SiO
molecular bands in the spectra of sources used as calibrators by the SWS.
Although OLP 10.0 was intended to be the final data release, these findings
have led to an additional release addressing this issue, OLP 10.1, which
corrects the artifacts.Comment: 14 pages, AASTex, including 5 figures. Accepted by ApJ Letter