Análisis del efecto de la porosidad y forma de un stent en un aneurisma sacular intracraneal utilizando el método Lattice Boltzmann

Abstract

This article presents an analysis of blood flow patterns in intracranial saccular aneurysm and the effects of the shape and porosity of the stents used in endovascular treatments. In this study will be used the flow reduction criteria for characterizing the efficiency of the stent. The hemodynamic properties of a newtonian blood flow into the aneurysm will be evaluated using the Lattice Boltzmann method (LBM). Porosity values and stent forms are proposed for analysis. In all stent cases analyzed is observed a reduction of velocity and pressure and an increase in viscosity. It is further noted that the rectangular contour stent is the optimal case and reduces the magnitude of the flow velocity inside the aneurysm much as 76%. The results help to understand the role of porosity in the form and design of a stent.PACS: 87.15.hj; 47.15.G-; 47.11.-jEn este artículo se presenta un analísis de los patrones de flujo sanguíneo en un aneurisma sacular intracraneal y los efectos de la forma y la porosidad de los stents empleados en tratamientos endovasculares. En este estudio se empleará el criterio de reducción del flujo para caracterizar la eficiencia del stent. Se evaluarán las propiedades hemodinámicas de un flujo sanguíneo newtoniano dentro del aneurisma a partir del método de Lattice Boltzmann (LBM). Se proponen algunos valores de porosidad y forma de stent para el análisis. En todos los casos de stent analizados se observa una reducción de velocidades y presiones y un aumento de viscosidad. Se observa además que el stent de contorno rectangular es el caso óptimo y reduce la magnitud de la velocidad del flujo en el interior del aneurisma hasta en un 76 %. Los resultados obtenidos ayudan a entender el papel de la forma y porosidad en el diseño de un stent.PACS: 87.15.hj; 47.15.G-; 47.11.-j

    Similar works