research

Repeatability of swept-source optical coherence tomography retinal and choroidal thickness measurements in neovascular age-related macular degeneration

Abstract

BACKGROUND: The aim was to determine the intrasession repeatability of swept-source optical coherence tomography (SS-OCT)-derived retinal and choroidal thickness measurements in eyes with neovascular age-related macular degeneration (nAMD). METHODS: A prospective study consisting of patients with active nAMD enrolled in the Distance of Choroid Study at Moorfields Eye Hospital, London. Patients underwent three 12×9 mm macular raster scans using the deep range imaging (DRI) OCT-1 SS-OCT (Topcon) device in a single imaging session. Retinal and choroidal thicknesses were calculated for the ETDRS macular subfields. Repeatability was calculated according to methods described by Bland and Altman. RESULTS: 39 eyes of 39 patients with nAMD were included with a mean (±SD) age of 73.9 (±7.2) years. The mean (±SD) retinal thickness of the central macular subfield was 225.7 μm (±12.4 μm). The repeatability this subfield, expressed as a percentage of the mean central macular subfield thickness, was 23.2%. The percentage repeatability of the other macular subfields ranged from 13.2% to 28.7%. The intrasession coefficient of repeatability of choroidal thickness of the central macular subfield was 57.2 μm with a mean choroidal thickness (±SD) of 181 μm (±15.8 μm). CONCLUSIONS: This study suggests that a change >23.2% of retinal thickness and 57.2 μm choroidal thickness in the central macular subfield is required to distinguish true clinical change from measurement variability when using the DRI OCT-1 device to manage patients with nAMD