Purpose Cadaveric study at our institution has demonstrated that optimal basaplate fixation of a reversed shoulder arthroplasty (RSA) could be achieved with screws in three major columns. Our aim was to review our early rate of aseptic glenoid loosening in a series of baseplate fixed according to this principle. Material and Methods Between 2005 and 2008, 48 RSA (Aequalis Reversed) were implanted in 48 patients with an average age of 74.4 years (range, 56 to 86 years). There were 37 women and 11 men. Twenty-seven primary RSAs were performed for cuff tear arthropathy, 3 after failed rotator cuff surgery, 6 for failed arthroplasties, 7 for acute fractures and 5 after failed ORIF. All baseplate fixation were done using a nonlocking posterior screw in the spine, a nonlocking anterior screw in the glenoid body, a locking superior screw in the coracoid and a locking inferior screw in the pillar. All patients were reviewed with standardized radiographs. The number of screws were reported. We measured the position of the screws in relation to the scapular spine and the coracoid process in two different views. We defined screw positions as totally, partially or out of the target. Finally we reported glenoid aseptic loosening which was defined as implant subsidence. Results Four patients were lost to follow-up. Thus, 44 shoulders could be reviewed after a mean follow-up of 13 months (range, 6 to 32 months). All baseplates were fixed with 4 screws. Thirty-seven (84%) screws were either partially or totally in the spine. Thus, 7 (16%) scapular spine screws were out of the target. No coracoid screw was out the target. Two (4.5%) patients had glenoid loosening. Both had a scapular spine and a coracoid screw partially in the bone. Conclusion Early aseptic glenoid loosening occurred before the two years follow-up and is most of time related to technical problems and/or insufficient bone stock and bone quality. Our study demonstrate that baseplate fixation according to the three columns principle is a reproducible technique and a valuable way to prevent early glenoid loosening

    Similar works

    Full text

    thumbnail-image