ICP MS ニヨル カンキョウ シリョウチュウ ノ コンセキ ゲンソ ノ ソクテイ 1 ICP MS ソウチ ノ サイテキ ジョウケン オヨビ ソノ ブンセキ セイド

Abstract

For determination of trace elements in environmental sample by inductively coupled plasma mass spectrometry (ICP-MS), the optimum condition of ICP-MS was determined and the stability of ICP-MS and the precision of elements concentration were investigated. The optimum condition was the sampling depth : 13 mm and RF-power : 1.2 KW. The stability of ICP-MS was indicated little drift of ionic strength after plasma lighting, however, ICP-MS was stabilized after 40 minutes. The coefficient variation (CV) of analyzed element concentration by external calibration was under 3% and the precision of analysis was very high

    Similar works