The CUORE experiment will search for neutrinoless double beta decay (0nDBD)
of 130Te using an array of 988 TeO_2 bolometers operated at 10 mK in the
Laboratori Nazionali del Gran Sasso (Italy). The detector is housed in a large
cryogen-free cryostat cooled by pulse tubes and a high-power dilution
refrigerator. The TeO_2 bolometers measure the event energies, and a precise
and reliable energy calibration is critical for the successful identification
of candidate 0nDBD and background events. The detector calibration system under
development is based on the insertion of 12 gamma-sources that are able to move
under their own weight through a set of guide tubes that route them from
deployment boxes on the 300K flange down into position in the detector region
inside the cryostat. The CUORE experiment poses stringent requirements on the
maximum heat load on the cryostat, material radiopurity, contamination risk and
the ability to fully retract the sources during normal data taking. Together
with the integration into a unique cryostat, this requires careful design and
unconventional solutions. We present the design, challenges, and expected
performance of this low-temperature energy calibration system.Comment: To be published in the proceedings of the 13th International Workshop
on Low Temperature Detectors (LTD), Stanford, CA, July 20-24, 200