Beta-gamma discriminator circuit

Abstract

The major difficulty encountered in the determination of beta-ray dose in field conditions is generally the presence of a relatively high gamma-ray component. Conventional dosimetry instruments use a shield on the detector to estimate the gamma-ray component in comparison with the beta-ray component. More accurate dosimetry information can be obtained from the measured beta spectrum itself. At Los Alamos, a detector and discriminator circuit suitable for use in a portable spectrometer have been developed. This instrument will discriminate between gammas and betas in a mixed field. The portable package includes a 256-channel MCA which can be programmed to give a variety of outputs, including a spectral display, and may be programmed to read dose directly