Safety of the Solitaire 4 × 40 mm Stent Retriever in the Treatment of Ischemic Stroke.

Abstract

PURPOSE Stent retrievers apply mechanical force to the intracranial vasculature. Our aim was to evaluate the safety and efficacy of the long Solitaire 4 × 40 mm stent retriever for large vessel occlusion in stroke patients. METHODS We conducted a retrospective analysis of all patients treated for acute ischemic large vessel occlusion stroke with the Solitaire 2 FR 4 × 40 device between May and October 2016 at our institution. Patient-specific data at baseline and at discharge were documented. Reperfusion was graded with the thrombolysis in cerebral infarction (TICI) classification. Postinterventional angiograms and follow-up cross-sectional imaging were used to evaluate complications. RESULTS TICI 2b/3 recanalization was achieved in 20 of 23 patients (87.0%), in 17 patients with the first retriever pass. NIHSS improved from a mean score at presentation of 16 (range 4-36) to 11 (range 0-41) at discharge. Mean mRS score at discharge was 3 (range 0-6) and 3 (range 0-6) at 90 days post-treatment. No infarcts in other territories were observed. One patient showed a (reversible) vasospasm in the postinterventional angiogram and another a small contrast extravasation in follow-up imaging. CONCLUSION The Solitaire 2 FR 4 × 40 stent retriever is a safe and efficient device for large vessel occlusion acute ischemic stroke with a high recanalization rate and a low peri- and postinterventional complication rate together with a good clinical outcome. Despite potentially higher friction and shearing forces, no increased incidence of visible damage to the vessel wall was observed