'Institute of Electrical and Electronics Engineers (IEEE)'
Doi
Abstract
This paper details the design of a closed-loop insulin delivery device, consisting of a glucose sensing circuit, and a basic microprocessor-based syringe pump. The glucose sensing circuit contains the required components to interface with CGMS\u27s glucose sensor assembly, while the syringe pump design uses microprocessor to allow flexible control over the pump driver. Instrumentation developed in this paper provides a ready reference to other researchers on the construction of a closed-loop insulin delivery apparatus with amperometric glucose sensor.<br /