HPLC-Electrospray Tandem Mass Spectrometry for Rapid Determination of Dihydropyrimidine Dehydrogenase Activity

Abstract

BACKGROUND: Patients with a partial dihydropyrimidine dehydrogenase (DPD) deficiency have an increased risk of developing severe 5-fluorouracil-associated toxicity. We developed a rapid and specific method to measure the DPD activity in peripheral blood mononuclear cells using HPLC tandem-mass spectrometry (HPLC-MS/MS). METHODS: The activity of DPD was measured with thymine as the substrate, followed by reversed-phase HPLC combined with electrospray ionization MS/MS and detection of the product dihydrothymine with multiple-reaction monitoring. Stable-isotope labeled dihydrothymine was used as the internal standard. RESULTS: Dihydrothymine was measured within an analytical run of 10 min, with a lower limit of quantification of 54 microg/L (0.4 micromol/L). The intraassay and interassay variations of the DPD activity assay were both <7%. A linear correlation (R(2) = 0.980; P <0.001) was observed between the HPLC-MS/MS data and those obtained with a reference method using radiolabeled thymine. There were no systematic differences between the 2 methods, and both methods yielded similar results. CONCLUSION: The analysis of the DPD activity with HPLC-MS/MS is rapid, accurate, and sufficiently sensitive to be used as a screening method for patients with a DPD deficienc

    Similar works

    Full text

    thumbnail-image

    Available Versions

    Last time updated on 01/04/2019
    Last time updated on 16/12/2017