We report the preparation of neutron-activated xenon for the calibration of
liquid xenon (LXe) detectors. Gamma rays from the decay of xenon metastable
states, produced by fast neutron activation, were detected and their activities
measured in a LXe scintillation detector. Following a five-day activation of
natural xenon gas with a Cf-252 (4 x 10^5 n/s) source, the activities of two
gamma ray lines at 164 keV and 236 keV, from Xe-131m and Xe-129m metastable
states, were measured at about 95 and 130 Bq/kg, respectively. We also observed
three additional lines at 35 keV, 100 keV and 275 keV, which decay away within
a few days. No long-lifetime activity was observed after the neutron
activation.Comment: to be published in NIM A, corrected typos in Table 1 and Fig.6 of the