The absolute calibration and characterization of the Multiband Imaging
Photometer for Spitzer (MIPS) 70 micron coarse- and fine-scale imaging modes
are presented based on over 2.5 years of observations. Accurate photometry
(especially for faint sources) requires two simple processing steps beyond the
standard data reduction to remove long-term detector transients. Point spread
function (PSF) fitting photometry is found to give more accurate flux densities
than aperture photometry. Based on the PSF fitting photometry, the calibration
factor shows no strong trend with flux density, background, spectral type,
exposure time, or time since anneals. The coarse-scale calibration sample
includes observations of stars with flux densities from 22 mJy to 17 Jy, on
backgrounds from 4 to 26 MJy sr^-1, and with spectral types from B to M. The
coarse-scale calibration is 702 +/- 35 MJy sr^-1 MIPS70^-1 (5% uncertainty) and
is based on measurements of 66 stars. The instrumental units of the MIPS 70
micron coarse- and fine-scale imaging modes are called MIPS70 and MIPS70F,
respectively. The photometric repeatability is calculated to be 4.5% from two
stars measured during every MIPS campaign and includes variations on all time
scales probed. The preliminary fine-scale calibration factor is 2894 +/- 294
MJy sr^-1 MIPS70F^-1 (10% uncertainty) based on 10 stars. The uncertainty in
the coarse- and fine-scale calibration factors are dominated by the 4.5%
photometric repeatability and the small sample size, respectively. The 5-sigma,
500 s sensitivity of the coarse-scale observations is 6-8 mJy. This work shows
that the MIPS 70 micron array produces accurate, well calibrated photometry and
validates the MIPS 70 micron operating strategy, especially the use of frequent
stimulator flashes to track the changing responsivities of the Ge:Ga detectors.Comment: 19 pages, PASP, in pres