In this paper, a first approach to the design of a portable device for non-contact monitoring
of respiratory rate by capacitive sensing is presented. The sensing system is integrated into a smart
vest for an untethered, low-cost and comfortable breathing monitoring of Chronic Obstructive
Pulmonary Disease (COPD) patients during the rest period between respiratory rehabilitation
exercises at home. To provide an extensible solution to the remote monitoring using this sensor and
other devices, the design and preliminary development of an e-Health platform based on the Internet
of Medical Things (IoMT) paradigm is also presented. In order to validate the proposed solution,
two quasi-experimental studies have been developed, comparing the estimations with respect to the
golden standard. In a first study with healthy subjects, the mean value of the respiratory rate error,
the standard deviation of the error and the correlation coefficient were 0.01 breaths per minute (bpm),
0.97 bpm and 0.995 (p < 0.00001), respectively. In a second study with COPD patients, the values
were -0.14 bpm, 0.28 bpm and 0.9988 (p < 0.0000001), respectively. The results for the rest period
show the technical and functional feasibility of the prototype and serve as a preliminary validation of
the device for respiratory rate monitoring of patients with COPD.Ministerio de Ciencia e Innovación PI15/00306Ministerio de Ciencia e Innovación DTS15/00195Junta de Andalucía PI-0010-2013Junta de Andalucía PI-0041-2014Junta de Andalucía PIN-0394-201