One-Step Detection of Pathogens and Viruses: Combining Magnetic Relaxation Switching and Magnetic Separation

Abstract

We report a sensing methodology that combines magnetic separation (MS) and magnetic relaxation switching (MS-MRS) for one-step detection of bacteria and viruses with high sensitivity and reproducibility. We first employ a magnetic field of 0.01 T to separate the magnetic beads of large size (250 nm in diameter) from those of small size (30 nm in diameter) and use the transverse relaxation time (<i>T</i><sub>2</sub>) of the water molecules around the 30 nm magnetic beads (MB<sub>30</sub>) as the signal readout of the immunoassay. An MS-MRS sensor integrates target enrichment, extraction, and detection into one step, and the entire immunoassay can be completed within 30 min. Compared with a traditional MRS sensor, an MS-MRS sensor shows enhanced sensitivity, better reproducibility, and convenient operation, thus providing a promising platform for point-of-care testing

    Similar works

    Full text

    thumbnail-image

    Available Versions