Abstract

<div><p>Background</p><p>Thoracoabdominal asynchrony is the nonparallel motion of the ribcage and abdomen. It is estimated by using respiratory inductive plethysmography and, recently, using optoelectronic plethysmography; however the agreement of measurements between these 2 techniques is unknown. Therefore, the present study compared respiratory inductive plethysmography with optoelectronic plethysmography for measuring thoracoabdominal asynchrony to see if the measurements were similar or different.</p><p>Methods</p><p>27 individuals (9 healthy subjects, 9 patients with interstitial lung disease, and 9 with chronic obstructive pulmonary disease performed 2 cycle ergometer tests with respiratory inductive plethysmography or optoelectronic plethysmography in a random order. Thoracoabdominal asynchrony was evaluated at rest, and at 50% and 75% of maximal workload between the superior ribcage and abdomen using a phase angle.</p><p>Results</p><p>Thoracoabdominal asynchrony values were very similar in both approaches not only at rest but also with exercise, with no statistical difference. There was a good correlation between the methods and the Phase angle values were within the limits of agreement in the Bland-Altman analysis.</p><p>Conclusion</p><p>Thoracoabdominal asynchrony measured by optoelectronic plethysmography and respiratory inductive plethysmography results in similar values and has a satisfactory agreement at rest and even for different exercise intensities in these groups.</p></div