Ag/AgCl chloride sensors were fabricated using thick-film technology. A number of different formulations were prepared and chloride responses were investigated over time. Near Nernstian, identical responses were observed over the first 160 days with an average chloride sensitivity of -51.8 ± 0.4 mV per decade change in chloride concentration (pCl), irrespective of paste formulation. After 6- months continuous immersion in tap water, pastes formulated with a glass binder began to exhibit a loss in sensitivity whilst those formulated from a commercial thickfilm dielectric paste remained functional for the best part of a year. The difference is attributed to the inclusion of proprietary additives in the commercial paste aiding adhesion and minimising AgCl leaching