An investigation was undertaken to make the elctrocardiography (ECG) and the electromyography (EMG) signal conditioning circuits two-fault tolerant and to update the circuitry. The present signal conditioning circuits provide at least one level of subject protection against electrical shock hazard but at a level of 100 micro-A (for voltages of up to 200 V). However, it is necessary to provide catastrophic fault tolerance protection for the astronauts and to provide protection at a current level of less that 100 micro-A. For this study, protection at the 10 micro-A level was sought. This is the generally accepted value below which no possibility of microshock exists. Only the possibility of macroshock exists in the case of the signal conditioners. However, this extra amount of protection is desirable. The initial part deals with current limiter circuits followed by an investigation into the signal conditioner specifications and circuit design