We deal with estimation of multiple dipoles from combined MEG and EEG
time--series. We use a sequential Monte Carlo algorithm to characterize the
posterior distribution of the number of dipoles and their locations. By
considering three test cases, we show that using the combined data the method
can localize sources that are not easily (or not at all) visible with either of
the two individual data alone. In addition, the posterior distribution from
combined data exhibits a lower variance, i.e. lower uncertainty, than the
posterior from single device.Comment: 4 pages, 3 figures -- conference paper from EMBEC 2017, Tampere,
Finlan