Superparamagnetic iron-oxide nanoparticles can be used in a variety of
medical applications like vascular or targeted imaging. Magnetic particle
imaging (MPI) is a promising tomographic imaging technique that allows
visualizing the 3D nanoparticle distribution concentration in a non-invasive
manner. The two main strengths of MPI are high temporal resolution and high
sensitivity. While the first has been proven in the assessment of dynamic
processes like cardiac imaging, it is unknown how far the detection limit of
MPI can be lowered. Within this work, we will present a highly sensitive
gradiometric receive-coil unit combined with a noise-matching network tailored
for the measurement of mice. The setup is capable of detecting 5 ng of iron in
vitro at 2.14 sec acquisition time. In terms of iron concentration we are able
to detect 156 {\mu}g/L marking the lowest value that has been reported for an
MPI scanner so far. In vivo MPI mouse images of a 512 ng bolus at 21.5 ms
acquisition time allow for capturing the flow of an intravenously injected
tracer through the heart of a mouse. Since it has been rather difficult to
compare detection limits across MPI publications we propose guidelines
improving the comparability of future MPI studies.Comment: 15 Pages, 7 Figures, V2: Changed the initials of Author Kannan M
Krishnan, added two citations, corrected typo