The response of AGATA segmented HPGe detectors to gamma rays in the energy
range 2-15 MeV was measured. The 15.1 MeV gamma rays were produced using the
reaction d(11B,ng)12C at Ebeam = 19.1 MeV, while gamma-rays between 2 to 9 MeV
were produced using an Am-Be-Fe radioactive source. The energy resolution and
linearity were studied and the energy-to-pulse-height conversion resulted to be
linear within 0.05%. Experimental interaction multiplicity distributions are
discussed and compared with the results of Geant4 simulations. It is shown that
the application of gamma-ray tracking allows a suppression of background
radiation following neutron capture by Ge nuclei. Finally the Doppler
correction for the 15.1 MeV gamma line, performed using the position
information extracted with Pulse-shape Analysis, is discussed.Comment: 10 pages, 11 figure