Near-infrared spectroscopy (NIRS)-derived measurement of oxidized cytochrome c oxidase concentration ([oxCCO]) has been used as an assessment of the adequacy of cerebral oxygen delivery. We report a case in which a reduction in conscious level was associated with a reduction in [oxCCO]. Hypoxaemia was induced in a 31-year-old, healthy male subject as part of an ongoing clinical study. Midway through the hypoxaemic challenge, the subject experienced an unexpected vasovagal event with bradycardia, hypotension and reduced cerebral blood flow (middle cerebral artery blood flow velocity decrease from 70 to 30 cm s(-1)) that induced a brief reduction in conscious level. An associated decrease in [oxCCO] was observed at 35 mm (-1.6 μM) but only minimal change (-0.1 μM) at 20-mm source-detector separation. A change in optical scattering was observed, but path length remained unchanged. This unexpected physiological event provides an unusual example of a severe reduction in cerebral oxygen delivery and is the first report correlating change in clinical status with changes in [oxCCO]