LUCAS Versus Manual Chest Compression During Ambulance Transport : A Hemodynamic Study in a Porcine Model of Cardiac Arrest

Abstract

Background-Mechanical chest compression (CC) is currently suggested to deliver sustained high-quality CC in a moving ambulance. This study compared the hemodynamic support provided by a mechanical piston device or manual CC during ambulance transport in a porcine model of cardiopulmonary resuscitation. Methods and Results-In a simulated urban ambulance transport, 16 pigs in cardiac arrest were randomized to 18 minutes of mechanical CC with the LUCAS (n=8) or manual CC (n=8). ECG, arterial and right atrial pressure, together with end-tidal CO2 and transthoracic impedance curve were continuously recorded. Arterial lactate was assessed during cardiopulmonary resuscitation and after resuscitation. During the initial 3 minutes of cardiopulmonary resuscitation, the ambulance was stationary, while then proceeded along a predefined itinerary. When the ambulance was stationary, CC-generated hemodynamics were equivalent in the 2 groups. However, during ambulance transport, arterial and coronary perfusion pressure, and end-tidal CO(2 )were significantly higher with mechanical CC compared with manual CC (coronary perfusion pressure: 43 +/- 4 versus 18 +/- 4 mmHg; end-tidal CO2: 31 +/- 2 versus 19 +/- 2 mmHg, P Conclusions-This model adds evidence in favor of the use of mechanical devices to provide ongoing high-quality CC and tissue perfusion during ambulance transport.Peer reviewe