Efficiency Determinations for a Ge(Li) Detector

Abstract

Author Institution: University of DaytonThe intrinsic efficiency of a Ge(Li) detector was determined for gamma-ray energies between 150 keV and 1500 keV. Two experimental methods were used. One method made use of the known relative intensities of La140 and Eu154 gamma rays. The second method made use of the known intensities of several calibrated sources. The resulting experimental full-energy peak-efficiency curve is compared with a published semi-empirical relation given by k « = [l-e-cr+Aoe-BE], c where e is the efficiency, and r and a are the photoelectric and Compton scattering absorption coefficients, respectively. The empirical constants k, A, B, and c are parameters, and E is the gamma-ray energy. It was found that the above relation agrees with the experimental results within an accuracy of 3%