In this paper we present a novel method for the correction of motion
artifacts that are present in fetal Magnetic Resonance Imaging (MRI) scans of
the whole uterus. Contrary to current slice-to-volume registration (SVR)
methods, requiring an inflexible anatomical enclosure of a single investigated
organ, the proposed patch-to-volume reconstruction (PVR) approach is able to
reconstruct a large field of view of non-rigidly deforming structures. It
relaxes rigid motion assumptions by introducing a specific amount of redundant
information that is exploited with parallelized patch-wise optimization,
super-resolution, and automatic outlier rejection. We further describe and
provide an efficient parallel implementation of PVR allowing its execution
within reasonable time on commercially available graphics processing units
(GPU), enabling its use in the clinical practice. We evaluate PVR's
computational overhead compared to standard methods and observe improved
reconstruction accuracy in presence of affine motion artifacts of approximately
30% compared to conventional SVR in synthetic experiments. Furthermore, we have
evaluated our method qualitatively and quantitatively on real fetal MRI data
subject to maternal breathing and sudden fetal movements. We evaluate
peak-signal-to-noise ratio (PSNR), structural similarity index (SSIM), and
cross correlation (CC) with respect to the originally acquired data and provide
a method for visual inspection of reconstruction uncertainty. With these
experiments we demonstrate successful application of PVR motion compensation to
the whole uterus, the human fetus, and the human placenta.Comment: 10 pages, 13 figures, submitted to IEEE Transactions on Medical
Imaging. v2: wadded funders acknowledgements to preprin