Novel
mass spectrometry (MS)-based proteomic tools with extremely
high sensitivity and high peak capacity are required for comprehensive
characterization of protein molecules in mass-limited samples. We
reported a nanoRPLC-CZE-MS/MS system for deep bottom-up proteomics
of low micrograms of human cell samples in previous work. In this
work, we improved the sensitivity of the nanoRPLC-CZE-MS/MS system
drastically via employing bovine serum albumin (BSA)-treated sample
vials, improving the nanoRPLC fraction collection procedure, and using
a short capillary for fast CZE separation. The improved nanoRPLC-CZE
produced a peak capacity of 8500 for peptide separation. The improved
system identified 6500 proteins from a MCF7 proteome digest starting
with only 500 ng of peptides using a Q-Exactive HF mass spectrometer.
The system produced a comparable number of protein identifications
(IDs) to our previous system and the two-dimensional (2D) nanoRPLC-MS/MS
system developed by Mann’s group with 10-fold and 4-fold less
sample consumption, respectively. We coupled the single-spot solid
phase sample preparation (SP3) method to the improved nanoRPLC-CZE-MS/MS
for bottom-up proteomics of 5000 HEK293T cells, resulting in 3689
protein IDs with the consumption of a peptide amount that corresponded
to only roughly 1000 cells